Settings
Oxi 7310
32
ba75931e03
12/2011
Setting the calibra-
tion interval
The calibration interval is set to 14 days in the factory.
You can change the interval (1 ... 999 days):
6.2
Sensor-independent settings
6.2.1
System
To open the
Storage & config
menu, press the
<F1_>
[Menu] key in the mea-
sured value display. After completing the settings, switch to the measured
value display with
<M>
.
Default settings are printed in
bold
.
1.
Open the menu for measurement settings with [Menu]
<F1>
.
2.
In the
Calibration
/
Calibration interval
menu, set the calibration interval
with
<
><
>
.
3.
Confirm the setting with
<ENTER>
.
4.
Quit the menu with
<M>
.
Menu item
Possible set-
ting
Description
System
/
General
/
Language
Deutsch
English
(further)
Selects the menu language
System
/
General
/
Beep
On
Off
Switches on/off the beep on key-
stroke
System
/
General
/
Illumination
Auto
On
Off
Switches the display illumination
on/off
System
/
General
/
Contrast
0 ...
50
... 100
%
Changes the display contrast
System
/
General
/
Switchoff time
10 min ...
1h
...
24 h
Adjusts the switch-off time
System
/
Interface
/
Baud rate
1200, 2400,
4800
, 9600,
19200
Baud rate of the data interface
System
/
Interface
/
Output format
ASCII
CSV
Output format for data transmis-
sion
For details, see section 8 T
System
/
Interface
/
Decimal separator
Dot (xx.x)
Comma (xx,x)
Decimal separator