Oxi 3310
What to do if...
49
ba75804e06
03/2012
6
What to do if...
Error message,
OFL
Error message,
Error
Sensor symbol flashes
Display
Meter does not react to
keystroke
You want to know which
software
version is in the meter
Cause
Remedy
– Measured value outside the
measuring range
– Use a suitable D.O. sensor
Cause
Remedy
– D.O. sensor contaminated
– Clean D.O. sensor and
replace it if necessary
Cause
Remedy
– Cleaning interval expired
– Recalibrate the measuring
system
Cause
Remedy
– Batteries almost empty
– Replace the batteries (see
section 5.1 M
AINTENANCE
)
Cause
Remedy
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press the <ENTER> and
<On/Off> key
simultaneously
Cause
Remedy
– E. g., a question by the service
department
– Switch on the meter.
Open the menu, <F1
_
>/
[Menü] / Speicher & Konfig. /
System / Service
Information. The instrument
data is displayed.