Oxi 315i
Resetting to default settings
7
Oxygen calibration procedure:
Resetting to default settings
Default
settings
The following settings are reset to the status on delivery when a reset is per-
formed (initialized):
Note
During a reset, the relative slope for the sensor types CellOx 325 and
DurOx 325 are reset together. They cannot be reset separately.
Reset procedure:
l
Keep a suitable air calibration vessel ready.
l
Connect the D. O. probe to the measuring instrument.
l
Put the D. O. probe into the air calibration vessel.
l
Press the <CAL> key. The oxygen calibration mode appears (display
shows O
2
CAL).
l
Press <RUN/ENTER>. The AutoRead measurement begins.
l
If the measured value is stable, the instrument displays the value of the
relative slope and the sensor evaluation.
l
Switch to the measuring mode with <M>.
6
7
1
A L
C
O
°
C
TP
Cal
6
7
1
8 8
0
O
°
C
TP
Cal
AR
S
l
Press <CAL> while pressing the <RUN/ENTER> key.
l
Press <RUN/ENTER> to reset the measurement parameters. The instru-
ment changes to the measuring mode.