Installation
DIQ/S 28X
42
ba77168d02
03/2017
Fig. 3-15 Prepared 24 V AC/DC line.
Connecting the
24 V AC/DC line
Fig. 3-16 Inserting the 24V AC/DC line
ca. 45 mm
wire 1
wire 2
4
On the left-hand side of the DIQ/S 28X, remove the two countersunk
screws and open the enclosure.
5
Screw a cable gland (pos. 1 in Fig. 3-16) with sealing ring (pos. 2) into
the enclosure below the 24 V AC/DC connection.
6
Loosen the coupling ring (pos. 3).
7
Feed the 24 V AC/DC line through the cable gland into the enclosure.
When doing so bend the flexible divider (pos. 4) to the right.
1
3
2
L
N
4