Installation
System 182
3 - 44
ba75503e07
04/2012
3.11.2 Connecting two sensors with compressed air cleaning
Connection scheme of
valve control
Example:
Relay 1 controls the cleaning of sensor 1.
Relay 2 controls the cleaning of sensor 2.
Variant: Relay 1 controls the cleaning of both sensors. Cleaning of both
sensors is carried out with the same settings.
Warning
In this configuration, the free switching contact (here: R3) may be
used to switch SELV voltages only.
Warning
No free wires must be allowed to project into the enclosure.
Otherwise, there is a danger that areas safe to contact could come
into contact with dangerous voltages. Always cut off any wires
that are not in use as closely as possible to the cable gland.
DIQ/CHV 1
SACIQ
SACIQ
Valve control line 1
Valve control line 2
Air
Air
DIQ/CHV 2
Sensor 2
Sensor 1
Max. total cable length
SNCIQ(/UG) and SACIQ = 250 m
DIQ/S 182-PR(/24V)
M
C
S
ESC
OK
Содержание DIQ/S 182-PR
Страница 6: ...Contents System 182 0 4 ba75503e07 04 2012 ...
Страница 10: ...Overview System 182 1 4 ba75503e07 04 2012 ...
Страница 14: ...Safety instructions System 182 2 4 ba75503e07 04 2012 ...
Страница 78: ...Operation System 182 4 18 ba75503e07 04 2012 ...
Страница 84: ...The SETTINGS menu System 182 5 6 ba75503e07 04 2012 ...
Страница 110: ...Relay outputs System 182 6 26 ba75503e07 04 2012 ...
Страница 122: ...PROFIBUS connection System 182 7 12 ba75503e07 04 2012 ...
Страница 124: ...Maintenance and cleaning System 182 8 2 ba75503e07 04 2012 ...
Страница 130: ...What to do if System 182 9 6 ba75503e07 04 2012 ...
Страница 142: ...Accessories and options System 182 11 2 ba75503e07 04 2012 ...
Страница 146: ...Appendix store separately if required System 182 13 2 ba75503e07 04 2012 ...