System 182
Installation
3 - 1
ba75503e07
04/2012
3
Installation
Note
How to connect the DIQ/S 182-PR to the PROFIBUS master is
described in detail in the chapter 7 PROFIBUS
3.1
Scope of delivery
The following parts are included in the scope of delivery of the DIQ/
S 182-PR:
DIQ/S 182-PR Universal Transmitter
Accessory kit with:
– Contact carrier with screws
– ISO cap nuts with screws and ring washers
– Cable glands with sealing gaskets
Operating manual.
3.2
Requirements of the measurement location
The measurement location must meet the environmental conditions
specified in section 10.1 DIQ/S 182-PR.
3.3
Installation guidelines for lightning protection
When using the IQ S
ENSOR
N
ET
instrumentation, particularly in outdoor
areas, adequate protection against (electrical) surges must be
provided. A surge is a summation phenomenon of surge voltage and
surge current. It is generated through the indirect effect of a lightning
event or switching operation in the mains, in the grounding system and
in information technology lines.
To be adequately protected against the damaging effects of surges, an
integrated concept of the following protective measures is required:
internal device-related protective measures and
external protective measures of the installation environment.
The internal device-related protective measures are already integrated
in the WTW online measuring technology as so-called 'lightning
protection' (see chapter 10 T
Содержание DIQ/S 182-PR
Страница 6: ...Contents System 182 0 4 ba75503e07 04 2012 ...
Страница 10: ...Overview System 182 1 4 ba75503e07 04 2012 ...
Страница 14: ...Safety instructions System 182 2 4 ba75503e07 04 2012 ...
Страница 78: ...Operation System 182 4 18 ba75503e07 04 2012 ...
Страница 84: ...The SETTINGS menu System 182 5 6 ba75503e07 04 2012 ...
Страница 110: ...Relay outputs System 182 6 26 ba75503e07 04 2012 ...
Страница 122: ...PROFIBUS connection System 182 7 12 ba75503e07 04 2012 ...
Страница 124: ...Maintenance and cleaning System 182 8 2 ba75503e07 04 2012 ...
Страница 130: ...What to do if System 182 9 6 ba75503e07 04 2012 ...
Страница 142: ...Accessories and options System 182 11 2 ba75503e07 04 2012 ...
Страница 146: ...Appendix store separately if required System 182 13 2 ba75503e07 04 2012 ...