MIQ/CR3; DIQ/CR3
Technical data
6 - 1
ba64107e08
05/2006
6
Technical data
Note
General technical data on MIQ modules are given in the T
ECHNICAL
DATA
chapter of the IQ S
ENSOR
N
ET
system operating manual.
Electrical data
Instrument safety
Terminal connections
Nominal voltage
Max. 24 VDC via the IQ S
ENSOR
N
ET
(for
details, see the T
ECHNICAL
DATA
chapter of
the IQ S
ENSOR
N
ET
system operating
manual)
Power consumption
3 W
Protective class
II
Overvoltage
category
II
Applicable norms
– EN 61010-1
– UL 3111-1
– CAN/CSA C22.2 No. 1010.1
IQ S
ENSOR
N
ET
connections
2
Additional connectable SENSORNET
terminator (terminating resistor)
Number of relay
contacts
3
Number of current
outputs
3
Terminal type
Screw-type terminal strip, accessible by
opening the lid
Terminal ranges
Solid wires:
0.2 ... 4.0 mm
2
AWG 24 ... 12
Flexible wires:
0.2 ... 2.5 mm
2
Line cross-section of
cables carrying mains
voltage
Europe:
1.5 ... 4.0 mm
2
USA:
AWG 14 ... 12
Cable feeds
4 cable glands M16 x 1.5 on the underside
of the module
Oxley Creek WWTP Water Reclamation (WTW - ba64107e08_xIQ_CR3) Vendor Manual
Doc Id: VM20
Owner: Howard Witten
Active Date: 15/12/2006