Settings
MIQ/CR3; DIQ/CR3
4 - 14
ba64107e08
05/2006
4.4
Deleting a link with an output
If a link from a current or relay output with a sensor is no longer
required, the link can be deleted.
The following steps describe operation with the 184 XT and 2020 XT
systems. With the System 182 XT, the outputs of the DIQ/CR3 can be
found directly under the
Settings
menu (see system operating manual
of the DIQ/S 182 XT).
Fig. 4-14 Settings of outputs and links:Erase link
1
Open the
Einstellungen/Settings
menu with
s
.
2
Using
d
and
g
, select and confirm the menu item,
System
settings
->
Settings of outputs and links
.
The
Settings of outputs and links
display opens.
3
Highlight the column
&
with
d
and confirm with
g
.
4
Highlight a linked output with
d
and confirm with
g
.
5
Select
Erase link
with
d
and confirm with
g
.
A security prompt appears.
6
Select
Erase link
with
d
and confirm with
g
.
The link is deleted.