Cond 3310 IDS
Settings
ba77051e04
07/2017
31
Setting menu
LR 925/01
Cell const. man
0.450 ...
0.475
...
0.500 cm
-1
Display and setting options for
the manually adjustable cell con-
stant.
This menu item is only available
when
Type
man
is set.
Temp. comp. (TC)
/
Method
nLF
Lin
off
Procedure for temperature com-
pensation (see section 5.2 T
26).
This setting is only available for
the measured parameters, con-
ductivity (
ϰ
) and resistivity (
ρ
).
Temp. comp. (TC)
/
Linear coeff.
0.000 ...
2.000
...
3.000 %/K
Coefficient of the linear tempera-
ture compensation.
This menu item is only available
when the linear temperature
compensation is set.
Temp. comp. (TC)
/
Reference temp.
20 °C
25 °C
Reference temperature
This setting is only available for
the measured parameters, con-
ductivity (
ϰ
) and resistivity (
ρ
).
Multiplier for TDS
0,40 ...
1,00
Multiplier for TDS value
Reset
-
Resets all sensor settings to the
delivery condition (see section
6.3.1 R
Menu item
Possible
setting
Explanation
Menu item
Possible
setting
Explanation
Cell constant
0.090 ...
0.100
...
0.110 cm
-1
Display and setting options for the
cell constant
Temp. comp. (TC)
/
Method
nLF
Lin
off
Procedure for temperature com-
pensation (see section 5.2 T
This setting is only available for the
measured parameters, conductivity
(
ϰ
) and resistivity (
ρ
).
Temp. comp. (TC)
/
Linear coeff.
0.000 ...
2.000
...
3.000 %/K
Coefficient of the linear tempera-
ture compensation.
This menu item is only available
when the linear temperature com-
pensation is set.