background image

PV830 Pneumatic PicoPump

World Precision Instruments 

9

OPERATING INSTRUCTIONS

Front Panel

Fig. 10—PV830 Front Panel.

!

CAUTION

: The valves always pass some air. Do NOT apply too much 

pressure when turning the knobs. They will not reach zero while the 

PV830

 is 

running. Over-tightening the knobs will damage the regulators.

Pressure Adjustment–

Adjust the  pressure by turning the 

Regulator

 knobs located 

next to the pressure gauges. Turn the knob clockwise to increase the pressure, and 
counter-clockwise to  decrease it. Because the regulating mechanism is self-venting, 
pressure is automatically released when the regulator setting is decreased. Although 
the pressure gauges cannot be read with high accuracy (especially at pressure 
settings below 2PSI), very reproducible pressures may be obtained, because the 
pressure regulators have a 20-turn dial yielding good pressure resolution. 

The pressure regulating mechanism functions by continuously bleeding a small 
amount of gas. This bleed rate increases as the difference between the input 
pressure and the output pressure increases. Loss of gas may be minimized by 
decreasing the difference between the input and output pressures, but some 
decrease in regulation may be noticed if the input pressure is not at least 10% 
greater than the output pressure. 

!

CAUTION

: To conserve tank contents, make sure to turn off the gas supply 

pressure when the 

PV830

 is not in use.

Vent–

The 

vac/hold

 switch selects between venting the 

Eject

 pressure port to either 

the 

Hold

 pressure (hold) or to the 

vac/atm line

 (vac). If the vacuum line is not in use 

and the switch is set to 

vac

, the pressure port is vented to the atmosphere.

Vacuum Adjustment–

Vacuum is adjusted by turning the vacuum 

Regulator

 knob 

located next to the vacuum gauge. Turning the knob clockwise increases the 
vacuum (lower pressure), and counter-clockwise decreases the vacuum. Because 

Содержание PicoPump PV830

Страница 1: ...PV830 Pneumatic PicoPump World Precision Instruments INSTRUCTION MANUAL Serial No _____________________ 042712 www wpiinc com...

Страница 2: ......

Страница 3: ...cking 2 INSTRUMENT DESCRIPTION 3 Setup 5 Vacuum Input 6 Pressure Input 6 Eject Pressure Port front panel 6 Vacuum Port 8 OPERATING INSTRUCTIONS 9 Front Panel 9 Rear Panel 10 Techniques In Microinjecti...

Страница 4: ...iv World Precision Instruments...

Страница 5: ...ally useful alternative to electroionophoresis since it does not mandate the use of charged ions Two different positive pressures may be applied one for ejection at high pressure and a second lower pr...

Страница 6: ...any parts are missing at 941 371 1003 or customerservice wpiinc com Returns Do not return any goods to WPI without obtaining prior approval RMA required and instructions from WPI s Returns Department...

Страница 7: ...position of the Vac Hold toggle switch This feature allows the pressure ejection channel to have a mild sucking action when pressure is not being applied A manual pneumatic switch also allows the vacu...

Страница 8: ...or Depress the plastic collar to remove tubing Fig 5 The PicoPump Pressure and Vacuum ports are located on the rear panel of the PV830 To install the hard tubing push it into the port until the tubing...

Страница 9: ...asily while metal tubing may require more effort For a diagram of the setup see Fig 6 Air or Nitrogen Supply 150 psi max Vacuum Pump or Line Pressure Input Vacuum Input Hold Eject Port Micropipette Ho...

Страница 10: ...o a nitrogen regulator on a N2 tank Recommended gases are dry air nitrogen or argon Never use corrosive gases If an air pressure line containing oil or water vapor is used an external filter is recomm...

Страница 11: ...hen using a MPH6S handle a firmer hold on the glass can be achieved by using two gaskets in the micropipette holder NOTE Capillary holders can be obtained from WPI which contain Ag AgCl half cells The...

Страница 12: ...ak when attaching or removing the tubing Before inserting the pipette place a gasket of the correct size in the cap Gasket Color Green Black Red White Pipette Diameter mm 1 0 1 2 1 5 1 65 WARNING DANG...

Страница 13: ...n dial yielding good pressure resolution The pressure regulating mechanism functions by continuously bleeding a small amount of gas This bleed rate increases as the difference between the input pressu...

Страница 14: ...electronic timer controls the time duration that the pressure solenoid stays open and the three commands serve only to start the timer Long pulses in the timed mode may be aborted by pressing the Stop...

Страница 15: ...helps you to correctly use the Hold pressure The backfilling flow rate is determined by the pressure of capillary action and the tip size Since the tip size is often determined by the requirement of t...

Страница 16: ...s cleaning the glass Small amounts of dust or grease can easily clog micron sized tips Cleaning with chromic acid solutions before pulling the electrode is commonly performed but care must be taken to...

Страница 17: ...ation The same technique may be used though if the drop is kept submerged under oil Droplets may seem to disappear after emergence from the tip Sometimes this is due to creepage of the aqueous fluid b...

Страница 18: ...approximately 1mV for every 10 PSI of applied pressure Multibarrel Microinjection For injection with multibarrel micropipette the PolyFil multibarrel micropipette coupling kit can be purchased from W...

Страница 19: ...e 1 Kit MPH6S for 1 5 mm pipette 5 ft tubing assembly 5430 20 PicoNozzle 1 Kit MPH6S for 2 0 mm pipette 5 ft tubing assembly 5430 ALL PicoNozzle 2 Kit MPH6S Micropipette Holder specify 1 0 1 2 1 5 or...

Страница 20: ...port will push the diaphragm and return the O ring to the seat To introduce a high pressure 1 Remove the 1 4 hard pressure tubing from the back of the pump 2 Use pieces of tubing with 1 4 OD to make a...

Страница 21: ...ng the knobs Over tightening the knobs will damage the regulators Vacuum Vacuum Input rear 0 30 0 in Hg 101kPa Vacuum Output 0 29 9 in Hg 101kPa Lowest Regulated Vacuum 3 in water 0 7kPa Regulator Acc...

Страница 22: ...hysical Specifications Power 95 130 VAC or 190 260 VAC switch selectable single phase 50 60 Hz 20VA Temperature 10 C 50 F to 40 C 104 F Dimensions 3 5 x 17 x 9 5 in 89 x 432 x 242 mm Mountable in stan...

Страница 23: ...L 215 m 267 m 1nL 100 m 1 24 m 100pL 46 m 57 m 10pL 21 5 m 27 m 1pL 10 m 12 4 m Fig 13 This graph compares the dimensions of a cube and a sphere with the same internal volume Volume of a Micropipette...

Страница 24: ...ion sensitive microelectrodes J Neuroscience Methods 8 231 247 1983 Deyhimi F and J A Coles Rapid silylation of a glass surface Choice of reagent and effect of experimental parameters on hydrophobici...

Страница 25: ...PV830 Pneumatic PicoPump World Precision Instruments 21...

Страница 26: ...22 World Precision Instruments DECLARATION OF CONFORMITY...

Страница 27: ...indirect special or consequential arising from a failure of this product to operate in the manner desired by the user WPI shall not be liable for any damage to data or property that may be caused dir...

Страница 28: ...44 0 1462 424700 Fax 44 0 1462 424701 E mail wpiuk wpi europe com Germany Zossener Str 55 10961 Berlin Tel 030 6188845 Fax 030 6188670 E mail wpide wpi europe com China Hong Kong WPI Shanghai Trading...

Отзывы: