background image

PV830 Pneumatic PicoPump

World Precision Instruments 

11

Techniques In Microinjection

One of the more demanding tasks for which the PicoPump has been designed is the 
microinjection of fluid into cells. Difficulties encountered will not usually involve the 
mechanics of the PicoPump, but rather the manufacture of suitable micropipettes. 
Both care and steady hands are needed in making pipettes to take full advantage of 
the features of the PicoPump. Here we list several important things to keep in mind 
when working with the PicoPump.

Setting the Hold Pressure

The 

Hold

 pressure is used to counterbalance the capillary action of the fluid 

backfilling into the pipettes. When an empty pipette is inserted into the fluid, a 
meniscus can be seen to rise up from the capillary tip. 

Hold

 pressure may be set 

by adjusting the regulator setting (with the 

Hold/vent

 switch in the 

hold 

position) 

until the meniscus stops at the desired position. In many applications, colored dye 
or fluorescent dye is dissolved in the injection fluid. The capillary effect can be seen 
with the color change at the tip of the pipette. When the fluid flows into the pipette, 
the color of the tip becomes lighter. If the hold pressure is higher than the capillary 
pressure, the fluid oozes out of the pipette. The solution around the pipette will be 
colored. Adjusting the 

Hold

 pressure prevents this from happening. 

Understanding how capillary action causes the backfilling of the pipette helps you 
to correctly use the 

Hold

 pressure. The backfilling flow rate is determined by the 

pressure of capillary action and the tip size. Since the tip size is often determined 
by the requirement of the application, controlling the pressure of capillary action 
becomes the main option to eliminate backfilling. 

The pressure of capillary action 

is determined by the inner diameter of the glass capillary where the meniscus 
of air/liquid interface is located. It has nothing to do with the pipette tip size, a 
common misunderstanding.

If we assume the pipette tip is near a cylindrical shape, the pressure of capillary 
action can be described by the LaPlace equation:
P=4

γ

 cos

θ

/d

where 

γ

 is the surface tension, 

θ

 is the contact angle between the water and glass, 

and d is the inner diameter of the capillary where the meniscus is located. In most 
cases, we can assume the contact angle for glass and water is zero (unless the 
glass surface is treated). From this equation we can see that the capillary pressure 
can vary a thousand times when the meniscus is moved from a 0.5µm ID tip to 
the 0.5mm shank. The pressure at 0.5µm tip is about 80 PSI (in aqueous solution) 
while at the shank will be only 0.08 PSI. Using one regulator to counterbalance the 
pressure in such a large dynamic range is not practical. The hold regulator in this 
instrument is optimized to work in the 0.2–10 PSI range. 10 PSI can counterbalance 
a meniscus at the section of tip where the inner diameter is 4µm. In practice, this is 
the highest pressure ever needed. On the lower pressure end, it becomes difficult to 
exactly counterbalance the capillary pressure when the meniscus is at the shank of 

Содержание PicoPump PV830

Страница 1: ...PV830 Pneumatic PicoPump World Precision Instruments INSTRUCTION MANUAL Serial No _____________________ 042712 www wpiinc com...

Страница 2: ......

Страница 3: ...cking 2 INSTRUMENT DESCRIPTION 3 Setup 5 Vacuum Input 6 Pressure Input 6 Eject Pressure Port front panel 6 Vacuum Port 8 OPERATING INSTRUCTIONS 9 Front Panel 9 Rear Panel 10 Techniques In Microinjecti...

Страница 4: ...iv World Precision Instruments...

Страница 5: ...ally useful alternative to electroionophoresis since it does not mandate the use of charged ions Two different positive pressures may be applied one for ejection at high pressure and a second lower pr...

Страница 6: ...any parts are missing at 941 371 1003 or customerservice wpiinc com Returns Do not return any goods to WPI without obtaining prior approval RMA required and instructions from WPI s Returns Department...

Страница 7: ...position of the Vac Hold toggle switch This feature allows the pressure ejection channel to have a mild sucking action when pressure is not being applied A manual pneumatic switch also allows the vacu...

Страница 8: ...or Depress the plastic collar to remove tubing Fig 5 The PicoPump Pressure and Vacuum ports are located on the rear panel of the PV830 To install the hard tubing push it into the port until the tubing...

Страница 9: ...asily while metal tubing may require more effort For a diagram of the setup see Fig 6 Air or Nitrogen Supply 150 psi max Vacuum Pump or Line Pressure Input Vacuum Input Hold Eject Port Micropipette Ho...

Страница 10: ...o a nitrogen regulator on a N2 tank Recommended gases are dry air nitrogen or argon Never use corrosive gases If an air pressure line containing oil or water vapor is used an external filter is recomm...

Страница 11: ...hen using a MPH6S handle a firmer hold on the glass can be achieved by using two gaskets in the micropipette holder NOTE Capillary holders can be obtained from WPI which contain Ag AgCl half cells The...

Страница 12: ...ak when attaching or removing the tubing Before inserting the pipette place a gasket of the correct size in the cap Gasket Color Green Black Red White Pipette Diameter mm 1 0 1 2 1 5 1 65 WARNING DANG...

Страница 13: ...n dial yielding good pressure resolution The pressure regulating mechanism functions by continuously bleeding a small amount of gas This bleed rate increases as the difference between the input pressu...

Страница 14: ...electronic timer controls the time duration that the pressure solenoid stays open and the three commands serve only to start the timer Long pulses in the timed mode may be aborted by pressing the Stop...

Страница 15: ...helps you to correctly use the Hold pressure The backfilling flow rate is determined by the pressure of capillary action and the tip size Since the tip size is often determined by the requirement of t...

Страница 16: ...s cleaning the glass Small amounts of dust or grease can easily clog micron sized tips Cleaning with chromic acid solutions before pulling the electrode is commonly performed but care must be taken to...

Страница 17: ...ation The same technique may be used though if the drop is kept submerged under oil Droplets may seem to disappear after emergence from the tip Sometimes this is due to creepage of the aqueous fluid b...

Страница 18: ...approximately 1mV for every 10 PSI of applied pressure Multibarrel Microinjection For injection with multibarrel micropipette the PolyFil multibarrel micropipette coupling kit can be purchased from W...

Страница 19: ...e 1 Kit MPH6S for 1 5 mm pipette 5 ft tubing assembly 5430 20 PicoNozzle 1 Kit MPH6S for 2 0 mm pipette 5 ft tubing assembly 5430 ALL PicoNozzle 2 Kit MPH6S Micropipette Holder specify 1 0 1 2 1 5 or...

Страница 20: ...port will push the diaphragm and return the O ring to the seat To introduce a high pressure 1 Remove the 1 4 hard pressure tubing from the back of the pump 2 Use pieces of tubing with 1 4 OD to make a...

Страница 21: ...ng the knobs Over tightening the knobs will damage the regulators Vacuum Vacuum Input rear 0 30 0 in Hg 101kPa Vacuum Output 0 29 9 in Hg 101kPa Lowest Regulated Vacuum 3 in water 0 7kPa Regulator Acc...

Страница 22: ...hysical Specifications Power 95 130 VAC or 190 260 VAC switch selectable single phase 50 60 Hz 20VA Temperature 10 C 50 F to 40 C 104 F Dimensions 3 5 x 17 x 9 5 in 89 x 432 x 242 mm Mountable in stan...

Страница 23: ...L 215 m 267 m 1nL 100 m 1 24 m 100pL 46 m 57 m 10pL 21 5 m 27 m 1pL 10 m 12 4 m Fig 13 This graph compares the dimensions of a cube and a sphere with the same internal volume Volume of a Micropipette...

Страница 24: ...ion sensitive microelectrodes J Neuroscience Methods 8 231 247 1983 Deyhimi F and J A Coles Rapid silylation of a glass surface Choice of reagent and effect of experimental parameters on hydrophobici...

Страница 25: ...PV830 Pneumatic PicoPump World Precision Instruments 21...

Страница 26: ...22 World Precision Instruments DECLARATION OF CONFORMITY...

Страница 27: ...indirect special or consequential arising from a failure of this product to operate in the manner desired by the user WPI shall not be liable for any damage to data or property that may be caused dir...

Страница 28: ...44 0 1462 424700 Fax 44 0 1462 424701 E mail wpiuk wpi europe com Germany Zossener Str 55 10961 Berlin Tel 030 6188845 Fax 030 6188670 E mail wpide wpi europe com China Hong Kong WPI Shanghai Trading...

Отзывы: