EVOM
plug connector. The inner electrodes are small Ag/AgCl pellet voltage sensors. They
are referred to as voltage electrodes and are connected to the
V
1
and
V
2
pins of the
telephone-type plug connector.
Instrument Self-tests
It is recommended that
EVOM
be put through the three self-tests described below
before using it for the first time and then periodically thereafter. When there is a concern
that the meter or the electrode is not functioning properly, the following three testing
protocols may be used to confirm
EVOM
’s operating status.
TESTING THE METER
This procedure tests the both the voltage function circuitry as well as that of the
resistance.
1.
Turn the Mode Switch to R.
2.
Turn the Power Switch on.
3.
Press the TEST R button
NOTE:
Do not press the TEST button when taking measurements or you may obtain a
false resistance measurement.
The meter should read as follows:
Range Switch Position
Display Reading
2000
Ω
1000
Ω
± 10
Ω
(±1%)
20 K
Ω
1 k
Ω
If the meter does not show these values, consult the “Troubleshooting” section in this
manual.
TESTING THE ELECTRODE FOR RESISTANCE MEASUREMENTS
1.
Insert the telephone-type plug at the end of the flexible electrode cable into the
Electrode Port on the meter.
2.
Immerse the electrode tips in a vessel containing an electrolyte solution similar to
the experimental culture medium or in 0.1-0.15 M potassium chloride (KCl) or sodium
chloride (NaCl).
Do not use a culture insert.
3.
Turn the Mode Switch to “R”
4.
Turn the Power on.
5.
When the meter is in the R mode and neither the Test R nor Measure R switch is
WORLD PRECISION INSTRUMENTS 5