Improving the Accuracy and Repeatability of the System
For measuring low-resistance tissue culture, or for more precise measurement, the user
should consider using the
Endohm
chamber instead of
STX2
electrode. Although cells
must be transferred from their culture wells to the Endohm chamber for measurement,
more accurate measurement of membrane resistance can be achieved. By introducing
concentric pairs of electrodes above and beneath the membrane, the
Endohm
reduces
background resistance (with a blank cup inserted) from 150
Ω
to less than 5
Ω
(except
Endohm-6
, which is higher). The shape of the current electrodes allows a more
uniform current density to flow across the membrane. With fixed electrode geometry,
the variation of readings on the same sample is 1-2
Ω
as compared to 5-10% of the
total reading using the
STX2
electrodes. The
Endohm
, together with the
EVOM
, offer
the most accurate, convenient and economical solution for trans membrane electrical
resistance measurement.