Introduction

The Iso-DAM8A is a modular bioamplifer designed for use with the Iso-DAM8 chassis, 

which can accommodate up to 8 modules. The Iso-DAM8A input stage is differential 

and is isolated from the AC power system. For example, there is no electrical 

connection to the AC power circuit ground. Isolated pre-amplification has important 

advantages over non-isolated amplifiers:

• 

There is no significant current can flow from the preamplifier input terminals 

to ground. Therefore, input connectors are intrinsically safe and cannot 

cause electrical shock at the site of measurement.

• 

Problems caused by “ground loops” are eliminated. 

Electrodes or other signal sources may be connected directly to the Iso-DAM8A 

input connector, or an optional active headstage (WPI Part 

#74040)

 may be used. 

The headstage connects to the Iso-DAM8A through a 5-foot cable. It contains a 

high impedance differential amplifier in a small metal housing. Electrodes may be 

connected directly to the headstage, which can be positioned at the measurement 

site, minimizing electrical noise pickup and stray capacitance.
The Iso-DAM8A module contains an input DC offset adjustment, switch-selectable 

gain, switch-selectable high pass and low pass filters, a line frequency “notch filter” 

(which can be set to 50 or 60 Hz), and an output DC position adjustment.
The Iso-DAM8 chassis can be ordered for either 110-120VAC or 220-240VAC 

operation. The chassis can accommodate up to eight Iso-DAM8A bioamplifiers. A 

separate BNC connector is provided for the output of each channel, and all channel 

outputs are available at a common DB-25 connector. 

Notes and Warnings

CAUTION

: The inputs to the Iso-DAM8A and the headstage can be damaged 

by electrostatic discharge. We recommend that you momentarily touch the 

Iso-DAM8A chassis or any grounded metal object before connecting wire or 

electrodes to the inputs.

Unpacking

Upon receipt of this instrument, make a thorough inspection of the contents and 

check for possible damage. Missing cartons or obvious damage to cartons should be 

noted on the delivery receipt before signing. Concealed damage should be reported 

at once to the carrier and an inspection requested. Please read the section entitled 

“Claims and Returns” on page 15 of this manual. Please contact WPI Customer 

Service if any parts are missing at 941.371.1003 or [email protected].

Returns:

 Do not return any goods to WPI without obtaining prior approval (RMA 

# required) and instructions from WPI’s Returns Department. Goods returned 

(unauthorized) by collect freight may be refused. If a return shipment is necessary, 

use the original container, if possible. If the original container is not available, use a 

suitable substitute that is rigid and of adequate size. Wrap the instrument in paper or 

Содержание Bridge-8

Страница 1: ...WORLD PRECISION INSTRUMENTS ISDB ISO DAM 8A and BRIDGE8 Serial No _____________________ www wpiinc com INSTRUCTION MANUAL 060920...

Страница 2: ......

Страница 3: ...Output to Zero 3 Bridge Switch 3 Transducer Calibration 4 Example 4 General Amplification 4 Specifications 4 ISO DAM8A 6 Introduction 7 Notes and Warnings 7 Unpacking 7 Instrument Description 8 Noise...

Страница 4: ...ii World Precision Instruments...

Страница 5: ...on Fig 1 Up to eight rackmountable ISO DAM8A and Bridge8 modules can be combined in a single chassis Introduction Any combination of ISO DAM8A or Bridge 8 modules may be combined in the same instrumen...

Страница 6: ...horized by collect freight may be refused If a return shipment is necessary use the original container if possible If the original container is not available use a suitable substitute that is rigid an...

Страница 7: ...vel to nearly zero volts adjust the Offset Adjust knob above the Gain switch so that both lamps are lit together A significant increase in the sharpness or resolution of the zero adjustment may be ach...

Страница 8: ...he output per mm Hg would be approximately 5 mV per mm Hg Since the nominal rating of 5 V per volt is usually not exact the user must in the end calibrate the transducer with a known hydrostatic press...

Страница 9: ...arasota FL 34240 9258 USA As the manufacture distributor of the apparatus listed declare under sole responsibility that the product s WPI PN ISDB SERIES To which this declaration relates is are in con...

Страница 10: ...at can cause damage to equipment Please read these carefully This symbol indicates a WARNING Warnings alert you to actions that can cause personal injury or pose a physical threat Please read these ca...

Страница 11: ...set to 50 or 60 Hz and an output DC position adjustment The Iso DAM8 chassis can be ordered for either 110 120VAC or 220 240VAC operation The chassis can accommodate up to eight Iso DAM8A bioamplifier...

Страница 12: ...he preparation and electrode lead wires in a Faraday cage or shield to reduce mains induction to sufficiently low levels It may be beneficial to extinguish overhead fluorescent lighting as well Use of...

Страница 13: ...ust the Gain switch position stability This potentiometer is accessible through a small screwdriver hole on the front panel To adjust the DC balance connect the and inputs of the input connector or th...

Страница 14: ...gly necessary if high gains are to be used with DC amplification The toggle switch below the DC Offset knob selects the polarity of the offset voltage If offset adjustment is not required the toggle s...

Страница 15: ...rode i e TM33B01 EP2 Ag AgCl Half Cell Prep A B Central Ground 3294 Grounding Clip Fig 7 A typical single ended configuration is shown here The Iso DAM8A inputs can be connected in either differential...

Страница 16: ...than a few inches from the amplifier inputs Even the use of shielded wires and proper grounding may be insufficient in a high electrical noise environment particularly if high impedance electrodes ar...

Страница 17: ...erely distort some signals If a large DC offset is present in the setup the DC OFFSET control can be used to cancel it The setting of this control can be quite sensitive especially at high gain settin...

Страница 18: ...14 World Precision Instruments...

Страница 19: ...ecial or consequential arising from a failure of this product to operate in the manner desired by the user WPI shall not be liable for any damage to data or property that may be caused directly or ind...

Страница 20: ...Tel 49 0 6031 67708 0 Fax 49 0 6031 67708 80 E mail wpide wpi europe com China Hong Kong WPI Shanghai Trading Co Ltd Rm 27A Liangfeng Building No 8 Dongfang Road Pudong New District Shanghai PR China...

Отзывы: