background image

ATC2000

World Precision Instruments 

7

OPERATING INSTRUCTIONS

The temperature controller provides electrically quiet control over the heating plate, 

so that you can place the animal directly onto the heating plate.
1.  Connect the heating plate to the 

Plate

 connector. 

2.  Connect the rectal temperature probe to one of the Inputs on the 

ATC2000

. If it 

is an RTD temperature sensor, plug it into the 

RTD

 port; if it is a thermocouple 

sensor, plug it into the 

TC

 port.

3.  Turn on the 

ATC2000

 using the power switch on the back of the unit.

4.  Adjust the set point, if necessary. See “Setting Temperature Set Point” on page 9.
5. 

The unit always powers up in Normal mode with the heat source set as 

default heat source from the previous use

. Typically, the plate temperature is 

the control value used until it reaches the set point. Set the source to the heater 

plate, and allow the unit to run until the plate reaches the set point. (See “Setting 

Controlled Temperature” on page 8.) The controller’s PID algorithm compares 

the controlled temperature to the preset (set point) temperature to determine 

the correct heater power output.

6.  After the plate reaches the set point, position the animal and insert the rectal 

probe.

7.  If desired, set the temperature source to the sensor monitoring the animal’s tem-

perature. See “Setting Controlled Temperature” on page 8. 

Display

The main display window shows the set point and the mode of operation on the left 

side of the screen. On the right side, the control temperature and plate temperature 

are visible. The arrow next to the TC Probe indicates that the TC Probe is the 

controlled temperature.

Fig. 7—The home page display shows the set point, mode, control temperature and plate 

temperature.

TEMPERATURE DISPLAY

In the illustration above, the unit is controlling the thermocouple (TC) probe. In this 

case, the RTD probe temperature can be viewed by pressing the 

Display

 button. 

When the unit controls the RTD probe, that temperature displays on this screen. The 

OK next to a temperature indicates that the temperature falls within the parameters 

defined by the low and high alarm limits. If the temperature falls below the low limit, 

LO displays instead of OK. Likewise, if it exceeds the high limit, HI displays. 

If the message “ERROR HI” appears on the Temperature Display, the temperature 

probe is either disconnected or defective. 

Содержание ATC2000

Страница 1: ...Instrumenting scienti c ideas WORLD PRECISION INSTRUMENTS ATC2000 Small Animal Temperature Controller Serial No _____________________ Version 2 7 www wpiinc com INSTRUCTION MANUAL 082318...

Страница 2: ......

Страница 3: ...daptive Mode 10 Setting Adaptive Interval 11 Setting Temperature Offset 12 Changing Shutdown Temperature Set Point 12 PID Controller Parameters 13 Setting PID Parameters 13 Resetting Factory Defaults...

Страница 4: ...ii World Precision Instruments...

Страница 5: ...netic radiation This is essential in electrophysiological recordings which are very sensitive to electromagnetic interference The controller uses proportional integral and derivative PID technology in...

Страница 6: ...r has three modes of operation MODE on page 8 Normal mode uses the configured sensor RTD or TC to control Adaptive mode uses the temperature of the heated plate and the temperature of the subject to c...

Страница 7: ...n inspection requested Please read the section entitled Claims and Returns on page 19 of this manual Please contact WPI Customer Service if any parts are missing at 941 371 1003 or customerservice wpi...

Страница 8: ...s on page 13 The error is subtracted from the previous value of the error to determine the difference It is then multiplied by the differential term D that is set in the configuration menu See Setting...

Страница 9: ...ly Default PID parameters Type P Proportional Gain I Integral Gain D Differential Gain Thermocouple 3900 2 5 0 RTD 600 0 4 0 Plate s 10230 0 41 0 You can reset PID parameters to the default setting by...

Страница 10: ...sing the configuration menu Status Indicator LED This heartbeat LED blinks continuously to indicate that the unit is operational and the control loops are running Like a heartbeat monitor it indicates...

Страница 11: ...termine the correct heater power output 6 After the plate reaches the set point position the animal and insert the rectal probe 7 If desired set the temperature source to the sensor monitoring the ani...

Страница 12: ...omewhat slower the normal mode depending on the sampling rate used NOTE This mode cannot be enabled when the heater plate is selected as the temperature source Shutdown is a fail safe mode If the plat...

Страница 13: ...isable the audible alarm Adjust the three temperature offsets Setting Temperature Set Point 1 Press the Config button to access the Configuration menu The temperature set point displays The default is...

Страница 14: ...cause they would beep continuously until the controlled temperature stabilizes When the audible alarm is enabled the controller will beep if the controlled temperature falls outside the range set by t...

Страница 15: ...o 10 minutes the temperature will be sampled again in 4 minutes and 18 seconds Setting Adaptive Interval The sampling interval for the adaptive mode should be set long enough to allow the subject and...

Страница 16: ...rature defaults to the temperature sensor embedded in the heating plate The shutdown mode has its own shutdown temperature set point which is different than the temperature set point in normal mode By...

Страница 17: ...Up or Down button to change the value 5 When you are finished adjusting the PID values press the Display button to save them and return to the home page screen Resetting Factory Defaults To reset all...

Страница 18: ...robe stainless steel for locating inflammation arteries etc Also for dental use FLEXIBLE IMPLANTABLE PROBES Designed for high accuracy on extremely small specimens such as insects seeds etc Maximum in...

Страница 19: ...r to solve this issue Improve the thermal coupling from the heating plate to the animal Place a heat conductive conformal material like cotton wadding be tween the animal and the plate Adjust the PID...

Страница 20: ...16 World Precision Instruments DECLARATION OF CONFORMITY...

Страница 21: ...MPLANTABLE PROBES 14 IT 1E 14 IT 18 14 IT 23 14 M MT 29 1 14 MT D 14 N NEEDLE MICROPROBES 14 noise 1 P parameters 13 parts list 2 PID 1 7 13 15 power 15 Q quiet 1 R rectal probes 14 rectal temperature...

Страница 22: ...18 World Precision Instruments...

Страница 23: ...direct special or consequential arising from a failure of this product to operate in the manner desired by the user WPI shall not be liable for any damage to data or property that may be caused direct...

Страница 24: ...D 61169 Friedberg Hesson Germany Tel 49 0 6031 1602171 Fax 49 0 6031 1602180 E mail wpide wpi europe com China Hong Kong WPI Shanghai Trading Co Ltd Rm 25e No8 Dongfang Rd Pudong District Shanghai 20...

Отзывы: