ATC2000
World Precision Instruments
15
TROUBLESHOOTING
Issue
Possible Cause
Solution
No
power
Poor connection
Check the power cable connection.
ERROR displays
instead of a temperature
Poor probe connection
Disconnect and properly re-connect the probes
to the
ATC2000
.
Animal can
-
not reach set temperature
Insufficient heating
power
The maximum power output on the heating
device is 30W. In most situations, the power is
sufficient to keep the animal warm up to 40°C.
To minimize heating loss, insulate the bottom
of heating plate and cover the animal with a
small blanket to keep it warm.
Alarm is sounding
Poor transfer of heat
from the heating plate to
the animal. (The plate is
hot, but the animal has
not warmed up.) You may
use these solutions inde-
pendently or together to
solve this issue.
Improve the thermal coupling from the heating
plate to the animal. Place a heat-conductive,
conformal material (like cotton wadding) be-
tween the animal and the plate.
Adjust the PID parameter to slow the plate
heating process. To do this, decrease the inte-
gral (I) value.
Poor connection to heat-
ing plate and probe
Disconnect both the heating plate and probe
and reinsert them properly.
NOTE
: If you have a problem/issue with that falls outside the definitions of this
troubleshooting section, contact the WPI Technical Support team at 941.371.1003 or
SPECIFICATIONS
The
ATC2000
conforms to the following specifications:
Temperature Range
up to 45°C
Resolution
0.1°C
Accuracy
± 0.3°C
Rat Sensor
RTD, OD 2.0mm tube with 3.5mm ball head (Optional
mouse sensor is available)
Maximum DC Output
10V, 3A
Power
Universal AC Adapter 90–264V Input, [email protected] max.
output
Dimensions
20.9 x 8.8 x 27.6 cm (8.2 x 3.5 x 10.9 in.)
Weight
5 kg (11 lb.)