![Worthington CS100 Скачать руководство пользователя страница 18](http://html.mh-extra.com/html/worthington/cs100/cs100_operation-and-maintenance-instructions-manual_998134018.webp)
18
CS100
Control System
Operation
About
This provides information about the control including the firmware version, the control serial number and
contact information. This is accessed through the menu system.
Temperature
The temperature in the Cryostorage freezer is measured at the location of the thermocouple. The installation
location is different depending on the model of the freezer but usually the thermocouple is positioned level with
the top storage box. This may mean that temperatures displayed may be slightly warmer than the temperature
experienced by the samples or product stored in the freezer.
Validation
Some organizations require that equipment be validated periodically. If information is needed on the proper
techniques to validate this equipment, please contact your supplier.
Menu System
MENU
1. Thermocouple Enable........................... On/Off
2. Set Date/Time..................................... Hr – Min – Mon – Day - Yr
3. Low Supply Alarm................................ 30-120
4. High Temp Alarm................................. 0° to -196°
5. Temp Alarm Delay................................ 0-60
6. Audible Alarm Retrigger........................ 0 to 720
7. Remote Alarm Delay............................. 0 to 720
8. Display Brightness................................ 1 to 15
9. About................................................. K SERIES V1.1 Firmware
SER
12345
Serial
Number
Factory Defaults
• Thermocouple....................................... On
• Date/Time............................................ Factory Set for Eastern Standard Time
• Low Supply Alarm................................. 30 minutes
• High Temp Alarm.................................. -100° C
• Temp Alarm Delay................................. 0 minutes
• Audible Retrigger.................................. 10 minutes
• Remote Alarm Delay.............................. 30 minutes
• Display Brightness................................. 7