CS100
Control System
Operation
Control Setting Adjustments
Level
The standard sensor assembly that is installed on a freezer consists of a circuit board with thermistors installed.
The assembly has a fixed fill range of 3”. The level can be changed by either raising or lowering the circuit
board within the freezer.
If a wider range is needed between the start fill and stop fill thermistor, please contact your supplier.
Menu System
Some control settings can be changed through the menu system.
Enter the menu system by pressing:
Move down through the menu system by pressing:
Move up through the menu system by pressing:
Select a menu choice or lock in a setting by pressing:
Back out of the menu system by pressing:
When changing settings, single button presses will increment/decrement a value one step at a time. Pressing
and holding a button will allow for rapid change of a setting.
The control will continue to monitor all sensors and conditions while the user accesses the menu system. If no
activity is detected for 30 seconds, the control will automatically revert back to the main operational screen.
Thermocouple Enable
The temperature sensor (thermocouple) can be turned on / off through the menu system. If the thermocouple
is turned on, the temperature is displayed on the LED display and a high temperature alarm will occur if
temperature exceeds the user-defined high temperature alarm setting. If the temperature sensor is turned off, no
temperature is displayed and no high temperature alarm can occur.
Set Date/Time
The date and time can be set through the menu system. The date and time is used to provide a date stamp
for data that is collected within the control. Use the appropriate buttons to adjust the settings and lock in the
values.