5.1.2 Post-operative treatment
The post-operative treatment must be carried out immediately, no later
than 30 minutes after the completion of the operation. The steps are as
follows:
1. Remove the optical fiber from the Curing light Device,and rinse away
the dirt on the surface of product with pure water (or distilled water/
deionized water);
2. Dry the product with a clean, soft cloth and place it in a clean tray.
Notes
a) The water used here must be pure water, distilled water or deionized
water.
5.2 Preparation before cleaning
Steps
Tools: tray, soft brush, clean and dry soft cloth Remove optical fiber
from main unit and put it into the clean tray.
Use a clean soft brush to carefully brush the optical fiber until the dirt
on surface is not visible. Then use soft cloth to dry the optical fiber and
put them into a clean tray. The cleaning agent can be pure water, distilled
water or deionized water.
5.3 Cleaning
The cleaning should be performed no later than 24 hours after the
operation.
The cleaning can be divided into automated cleaning and manual
cleaning. Automated cleaning is preferred if conditions permit.
5.3.1 Automated cleaning
•The cleaner is proved to be valid by CE certificationin accordance with
ENISO 15883.
▪There should be a flushing connector connected to the inner cavity of
the product.
▪The cleaning procedure is suitable for the product, and the irrigating
period is sufficient.
It is recommended to use a washer-disinfector in accordance with EN
ISO15883. For the specific procedure, please refer to the automated
disinfection section in the next section “Disinfection”.