background image

ZMN-SM-318    V1.0- 20210118

Содержание i-Scan

Страница 1: ......

Страница 2: ...ons 11 4 Notes 19 5 Troubleshooting 20 6 Cleaning disinfection and sterilization 20 7 Storage maintenance and transportation 21 8 Environment protection 22 9 After sales service 22 10 Electromagnetic...

Страница 3: ...cking list 1 4 Structure and components This product is mainly composed of main unit IP imaging plate power adapter disposable protective bag of IP imaging plate protection card image management softw...

Страница 4: ...shed 1 8 5 Image uniformity 2 1 8 6 Effective imaging area Specifications Maximum dimensions mm X direction X direction 0 21 32 1 23 39 2 30 40 3 26 53 1 8 7 Class laser product complying with the EN...

Страница 5: ...x Be careful not to drop or damage the equipment 2 2 2 Power adapter installation Take out the power adapter and power cord from the packing box and connect them as shown in the figure Note Only the p...

Страница 6: ...er needs a WIFI module if the computer does not have a built in WIFI module an external WIFI module needs to be inserted Display resolution 1280 x 1024 or above 2 3 2 Dental X ray generator compatibil...

Страница 7: ...on do not agree cancel the installation as shown in Figure 5 Figure 5 e Select components The user selects the corresponding component as needed and then click the Next button as shown in Figure 6 Fig...

Страница 8: ...as shown in Figure 8 Figure 8 h After the Install button is clicked the program starts to install The user just waits for the completion of installation as shown in Figure 9 Figure 9 i After the soft...

Страница 9: ...9 Figure 10 2 3 4 WIFI connection a Click Settings on the home screen...

Страница 10: ...nect e Open the software click the Acquisition button and select the I SCAN device type After connecting the device the IP address will be displayed below the device type Click open and the device wil...

Страница 11: ...protective cover of IP imaging plate to avoid the risk of cross infection 3 1 3 Insert the IP imaging plate into the hard protection card in the disposable protective bag The black side of the IP imag...

Страница 12: ...ective gloves Disinfect and wash hands 3 2 5 Remove the adhesive tape and open the disposable protective bag of IP imaging plate 3 2 6 Remove the hard protection card containing the IP imaging plate f...

Страница 13: ...m or the image data of imaging plate scanner is not erased it is necessary to conduct erasing 3 5 Software operation instructions 3 5 1 Login module Double click Ai Dental Server to start the server A...

Страница 14: ...ication can be successful If you need to delete a patient click the Delete Patient button and click OK As is shown in Figure 14 Figure 14 b Image acquisition After entering the software interface the...

Страница 15: ...e Click the Diagnosis button to enter the diagnosis module Click the image on the left to select an image to be processed There are image processing tools on the right side of the diagnosis interface...

Страница 16: ...use button to form the starting point and move the mouse Click the left mouse button again to form the end point and right click the end point to end the measurement The measurement line will be displ...

Страница 17: ...the Sign Out button to return to the login interface Click the language drop down box to select the software language Click the tooth profile drop down box and select the tooth profile number As is s...

Страница 18: ...Figure 21 d IP setting Click the IP Setting button to enter the IP setting interface Enter the IP address and port number Click Test Connection to view the test result Click Modify to switch the conn...

Страница 19: ...tective bag necessary health protection measures must be taken to avoid the risk of cross infection 4 5 The imaging plate scanner complies with the IEC 60601 1 standard Only peripheral equipment confo...

Страница 20: ...ld be thoroughly disinfected before use after each repair or maintenance 6 1 Cleaning and disinfection of imaging plate scanner and its detachable parts a The imaging plate scanner and its detachable...

Страница 21: ...ents or corrosive cleaning products to disinfect the imaging plate scanner and IP imaging plate d Do not spray disinfectant directly on the imaging plate scanner 7 Storage maintenance and transportati...

Страница 22: ...romagnetic compatibility information specified in this manual Portable and mobile radio frequency communication equipment may affect this device The following cables must be used to meet electromagnet...

Страница 23: ...urposes Harmonic emission GB17625 1 Class A Voltage fluctuation flicker emission GB17625 2 Complies 10 2 Guidance and manufacturer s declaration electromagnetic immunity Guidance and manufacturer s de...

Страница 24: ...anner requires continued operation during power mains interruptions it is recommended that the scanner be powered from an uninterruptible power supply or a battery Power frequency magnetic field 50Hz...

Страница 25: ...Hz where P is the maximum output rated power of the transmitter provided by the transmitter manufacturer in watts W and d is the recommended separation distance in meters m Field strengths from fixed...

Страница 26: ...uld be less than 3V m 10 4 Recommended separation distances between portable and mobile RF communications equipment and the imaging plate scanner Recommended separation distances between portable and...

Страница 27: ...n is affected by the absorption and emission from buildings objects and human bodies Notes a Without the explicit consent of Woodpecker unauthorized changes or modifications to the equipment may cause...

Страница 28: ...perate the equipment This equipment should be installed and operated with a minimum distance of 20cm between the radiator and your body This equipment has been tested and found to comply with the limi...

Страница 29: ...29 Connect the equipment into an outlet on a circuit different from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for help...

Страница 30: ...ZMN SM 318 V1 0 20210118...

Отзывы: