Section 13 Troubleshooting
Display
Cause
Solution
E-1
The sensor area is damaged
or contaminated
Ensure the sensor area is clean and that there
are no objects covering the sensor area. Ensure
no used test strips are left in the holder. Refer
to the ‘Maintenance’ section. Restart the
meter. Contact Woodley Equipment Company
if the sensor area window is broken.
E-2
Test strip was removed
during testing
Repeat the test and ensure the test strip
remains in place during testing.
E-3
Test strip expired
Replace test strips and code chip. Check their
expiry date.
E-4
Battery power is low. The
meter cannot be tested
until it is charged
Connect the meter to PC via USB to charge.
E-6
Code chip was removed
Insert correct code chip. Confirm the code chip
matches the test strip code and repeat the test.
E-7
The temperature is out of
the operating range
Ensure the meter is in an environment where
the temperature is between 10-35°C.
HI
The result is higher than the
readable range
Check sampling technique and repeat test with
a new test strip. Run a quality control check.
Contact Woodley Equipment Company.
Lo
The result is lower than the
readable range