19
Section 8 Control Strip Test
Press and hold
to enter the menu, use
◄
or
►
to scroll across to CTR. Press
to enter
CTR Mode. After entering the CTR Mode, the meter will display OPt. Press
and then
insert a control strip into the Test Strip Holder. Follow the direction of the arrows indicated
on the strip. Ensure that the control strip is inserted all the way. The optical check will start
automatically. If the meter displays
‘
OH
’
, the meter is normal. If the meter displays
‘
FL
’
, the
meter is not functioning properly.
Note:
•
The control strip is intended to check the optical system
•
Allow the control strips and the meter to reach operating temperature (10-35°C)
prior to testing
•
The optical check should be performed under normal lighting conditions. Do not
perform under direct sunlight or extreme lighting conditions
If the meter displays
‘
FL
’
, check the control strip for contamination or check if it is damaged.
If there are any visible signs of damage or contamination, discard the control strip and retest
using a new control strip.
Once the control strip is removed, the meter will return to the Initial Screen.