Section 9 Quality Control 

 

Each laboratory should have its own protocol for running quality control. 
 
A quality control should be performed: 

 

When a new package of test strips is opened 

 

When a new operator uses the meter 

 

When test results seem inaccurate 

 

After performing maintenance or service on the meter 

 
If quality control results are outside of the range, perform the following checks: 

 

Ensure that the test strips used have not expired 

 

Ensure that the test strips have been stored in the vial correctly 

 

Ensure that the controls have not expired 

 

Repeat the test to ensure no errors were made during the test 

 

Control Solution Testing 

Renal Function Control Solution testing is performed in a very similar manner to blood tests. 
The InSight Mini Chem Control Solution is used instead of blood. 
 

Note:

 Make sure the control solution and all the test materials reach operating 

temperatures of 10-35°C prior to testing. Tests can only be accurately performed when the 
control solutions and test materials are within this temperature range. 
 

1.

 

Turn on the meter 

2.

 

Insert the code chip into the meter. Refer to Section 4 

‘Coding the Meter’ 

for details 

3.

 

Compare the code number on the code chip with the code number printed on the 
test strip pouch label and ensure the two numbers are identical to avoid inaccurate 
results 

4.

 

Press and hold 

 for 2 seconds to enter the Settings mode and press 

 to change 

the selection. Press 

 to enter the CTR. Press 

 or 

 to switch between OPt and 

SOLU. Press 

 once SOLU is selected. 

5.

 

The meter will flash the test strip icon. Insert a test strip completely into the Test 
Strip Holder in the same direction as the arrows printed on the strip  

6.

 

When the meter is flashing the blood drop symbol, open the screw cap of the 
control solution bottle and turn the bottle upside down. Squeeze the control 
solution bottle gently and discard the first drop. If there are bubbles in the previous 
drop, squeeze the bottle and discard another drop until there are no bubbles in the 
drop. Apply the next drop to the specimen well on the test strip while keeping the 
bottle vertical. Make sure the control solution is applied directly into the specimen 
well and there are no bubbles in the solution drop. The meter will automatically 

Содержание InSight Mini Chem

Страница 1: ...User Manual...

Страница 2: ...odley Equipment Company may void the warranty of the InSight Mini Chem Do not use accessories which are not supplied or recommended by Woodley Equipment Company Do not use the meter if it is damaged P...

Страница 3: ...13 Section 6 Testing 14 Section 7 Memory Communication 18 Section 8 Control Strip Test 19 Section 9 Quality Control 20 Section 10 Maintenance 22 Section 11 Precautions 23 Section 12 Limitations 24 Sec...

Страница 4: ...ons carefully and complete any necessary training before use Use the code chip that is included in each box of test strips Do not mix code chips from different lots Only use the InSight Mini Chem Test...

Страница 5: ...lease contact Woodley Equipment Company Before testing read the instructions carefully and learn about all the components of the InSight Mini Chem Meter Depending on the packaging type some of the com...

Страница 6: ...lso eject the used lancets 6 Sterile Lancets Used with the lancing device to draw capillary blood samples Sterile lancets are inserted into the lancing device for each blood draw and are discarded aft...

Страница 7: ...ry and avoid exposing it to extreme temperatures and humidity Do not drop the meter or get it wet If the meter is dropped or has got wet ensure the meter is working properly by running a Control Check...

Страница 8: ...the arrows indicate the direction in which the test strip should be inserted into the meter Sample Collection Area After the test strip is inserted into the Test Strip Holder apply the correct sample...

Страница 9: ...trip If the E 2 error or another error refer to Section 13 for error codes appears on the display please discard the used strip and retest with a new strip Code Number Printed on each package of test...

Страница 10: ...en 2 30 C Avoid exposure to direct sunlight extreme temperatures and humidity Control strips should be stored in their tightly capped canister at room temperature to keep them in working condition Kee...

Страница 11: ...doesn t fall within the range refer to the InSight Mini Chem Control Solution Package Insert for further instructions Precautions Make sure the control solution and all the test materials reach an op...

Страница 12: ...meter Compare the code number on the code chip with the code number printed on the test strip packaging or the foil pouch Results may be inaccurate if the two numbers are not identical Insert the new...

Страница 13: ...ess or to change it Press to save Minute Setup As the minute is displayed press or to change it Press to save Sound Setup As the sound is displayed press or to switch between On and Off Press to save...

Страница 14: ...ecimen well before testing in order to ensure the cellular components are evenly distributed Allow the specimen to attain operating temperature 10 35 C for approximately 15 minutes if the specimen has...

Страница 15: ...in the lancet holder Hold the lancet firmly in the lancet holder and twist the safety tab of the lancet until it loosens Then pull the safety tab off the lancet Save the safety tab for lancet disposa...

Страница 16: ...tate and pull off the protective cap 2 After cleaning the skin hold the lancet firmly against the puncture site 3 Press the lancet against the puncture site Discard the lancet in an appropriate sharps...

Страница 17: ...en and is ready for another test strip to be inserted and to perform a new test Note Discard all blood samples used test strips and materials carefully Treat all blood samples as if they were infectio...

Страница 18: ...the data from the meter s memory press and at the same time to enable data deletion Press to return to the Setting screen Communication If Bluetooth is turned on Bluetooth icon is displayed when quer...

Страница 19: ...meter displays FL the meter is not functioning properly Note The control strip is intended to check the optical system Allow the control strips and the meter to reach operating temperature 10 35 C pr...

Страница 20: ...is temperature range 1 Turn on the meter 2 Insert the code chip into the meter Refer to Section 4 Coding the Meter for details 3 Compare the code number on the code chip with the code number printed o...

Страница 21: ...f the bottle and drop it freely into the sample well Avoid touching the strip with the tip of the bottle Interpreting Results The results should fall within the range s printed on the bottle label and...

Страница 22: ...getting liquids residue or control solutions in the meter through the Test Strip Holder Code Chip Slot or USB Port Charging Step The meter is powered by a lithium battery and can be charged by a USB p...

Страница 23: ...to Appendix 1 Meter Specifications Page 28 for operating condition requirements Keep the meter clean Wipe it frequently with a soft clean and dry cloth Use a damp cloth when needed Do not clean the me...

Страница 24: ...L Haemoglobin 2 g L 200 mg dL Ibuprofen 2425 mol L 50 mg dL Dopamine 5 87 mol L 0 09 mg dL Methyldopa 71 mol L 1 5 mg dL High concentrations of uric acid and ascorbic acid can lead to low measurements...

Страница 25: ...ting E 3 Test strip expired Replace test strips and code chip Check their expiry date E 4 Battery power is low The meter cannot be tested until it is charged Connect the meter to PC via USB to charge...

Страница 26: ...20 UR 0 90 10 00 mmol L 5 41 60 06 mg dL 2 00 mmol L 12 01 mg dL 10 01 40 00 mmol L 60 07 240 24 mg dL 20 Precision Item Range of Sample SD UA 90 300 mol L 1 51 5 04 mg dL SD 0 023 mmol L 0 38 mg dL...

Страница 27: ...venous whole blood plasma and serum Sample Volume 35 l Power Source USB to micro USB cable for charging meter Note During charging the meter is non operational Units of Measurement UA mol L CR mol L U...

Страница 28: ...must be consulted in all cases where this symbol is marked Symbol for the marking of electrical and electronic devices according to Directive 2012 19 EU The device accessories and packaging have to b...

Страница 29: ...eter is warrantied for a period of 12 months from date of invoice Woodley Equipment Company Ltd Old Station Park Buildings St John Street Horwich Bolton BL6 7NY UK Tel 44 0 1204 669033 Email sales woo...

Страница 30: ...30 Old Station Park Buildings St John Street Horwich Bolton BL6 7NY UK Tel 44 0 1204 669033 Email sales woodleyequipment com Web www woodleyequipment com...

Отзывы: