Background Illumination and Contrast

Activate the 

Display

 key.

In order to extend battery service life, display illumination can be 
switched off entirely. 
Press the corresponding softkey for this.

If LCD illumination is activated (= ON), it is automatically switched off 
several seconds after the last key has been activated in order to extend 
battery service life. As soon as a key is activated again, illumination is 
switched back on.

Contrast can be optimized with the two keys at the far right.

Press the 

START 

key to quit the setup menue. The data are saved to 

memory.

Setting the Clock

 

Activate the 

Time

 key.

The cursor appears at the first digit in the date. Enter the desired
number with one of the softkeys. 
Numbers which do not appear in the window can be displayed with 
the help of the 

 or the 

 

key. Each time a number is selected, 

the cursor moves to the next position to the right.

Date and time are saved as soon as the last number has been 
entered.

Press the 

START 

key to quit the setup menue. The data are saved to 

memory.

low

high

Contrast

off

on

LCD Illumination

Содержание METRISO 2000

Страница 1: ...METRISO 2000 ESD Test Instrument High Resistance Tester for the Measurement of Resistance to Ground and Surface Resistance 3 349 087 53 4 8 01 Operating Instructions...

Страница 2: ...functions Right switch contact Displays basic functions scrolling from right to left scrolling from left to right Battery level display finger contact and measurement input Indicator lamps menu driven...

Страница 3: ...tance 11 5 4 1 Point to Point Measurement RS with 2 Measuring Probes according to IEC 61340 4 1 DIN IEC 61340 4 1 11 5 4 2 Measuring Surface Resistance RS of Packaging Materials with a Ring Probe in c...

Страница 4: ...the METRISO 2000 Resistance to ground surface and volume resistance with special measuring probes Insulation resistance with variable and rising measurement voltage Temperature and relative humidity w...

Страница 5: ...can be used as well Be absolutely sure to refer to chapter 12 2 page 22 regarding the charging cycle and the charging adapter Always replace the batteries in complete sets Dispose of batteries in an...

Страница 6: ...is switched on Press the key to quit the setup menue Setting On Time Manual Shutdown Activate the Setup key Press the Ton key and then the 10sec 20sec 30sec or 60sec key depend ing upon the desired du...

Страница 7: ...back on Contrast can be optimized with the two keys at the far right Press the START key to quit the setup menue The data are saved to memory Setting the Clock Activate the Time key The cursor appears...

Страница 8: ...in digital format until the next measuring sequence is started or until the instrument switches itself off automatically If the measuring range upper value is exceeded the upper value is dis played pr...

Страница 9: ...ch exceed the selected limit value the red LIMIT LED lights up Select the limit value menu Set the limit value Setting the Limit Value Display the desired numbers and if required the decimal point wit...

Страница 10: ...Person to Ground RE System or RG System e g for the measurement of a resistance through the combination of a person shoes flooring system according to ESD STM 97 1 1999 and IEC 61340 4 5 NP Connect th...

Страница 11: ...volts The value then displayed is the result Test Voltage according to IEC 61340 5 1 DIN EN 61340 5 1 5 4 2 Measuring Surface Resistance RS of Packaging Materials with a Ring Probe in compliance with...

Страница 12: ...value resistance necessitates extremely small measuring current and may be rendered problematic by influences such as electromagnetic fields humidity or surface current A precise test setup is therefo...

Страница 13: ...ng Data Display the desired alphanumeric character by pressing the or the key and then select the character with the corresponding softkey Control characters can be entered in the same way and have th...

Страница 14: ...alphanumeric character with the or the key and select the desired character with the corresponding softkey Control characters are entered in the same way and have the following meanings reverse and de...

Страница 15: ...oon as an individual data record is deleted 6 3 2 Deleting a Memory Address Data Function Select Data Enter blanks to the data fields LOCATION Department Nr and INDEX After all of these fields have be...

Страница 16: ...Enter memory address 000 After acknowledgement by pressing button START press at center a safety request appears By simultaneously activating O and K all data that have been saved are deleted from mem...

Страница 17: ...en entered exit the data entry window by pressing the softkey If the selected voltage is not within the valid range the cursor jumps back to the entry position for the first digit 7 2 Tests with Risin...

Страница 18: ...he housing Then push the adapter down to fix it tightly Press the T R H key The adapter is activated via the interface Select the desired temperature unit of measure C or F with the corresponding key...

Страница 19: ...et of batteries alkaline manganese with 10 s on time and performance of one measurement before each automatic instrument shutdown For U f and T FREL 5 hours same conditions as above Measured Quantity...

Страница 20: ...gory 600 V CAT III Fouling Factor 2 Electromagnetic Compatibility EMC IEC 61326 EN 61326 Ambient Conditions Nominal Temperature 0 40 C Operating Temperature 10 50 C Storage Temperature 20 60 C without...

Страница 21: ...D 12 Maintenance 12 1 Self Test The self test is started from the main menu with the Test key The test has a duration of several seconds The following information is displayed in the two headers Type...

Страница 22: ...ries The instrument has been disconnected from the measuring circuit at all poles The voltage selector at the charger has been set to 9 V Connect the NA 0100S battery charger to the charging socket wi...

Страница 23: ...WOLFGANG WARMBIER 23...

Страница 24: ...nted in Germany Subject to change without notice WOLFGANG WARMBIER Untere Gie wiesen 21 78247 Hilzingen Germany Phone 49 77 31 86 88 0 Fax 49 77 31 86 88 30 e mail info warmbier com http www warmbier...

Отзывы: