12
WOLFGANG WARMBIER
➭
If the value is displayed with R
X
> 100 k
W
, the test voltage must be
changed to 100 volts. The value then displayed is the result.
Test Voltage according to IEC 61340-5-1 / DIN EN 61340-5-1:
5.5
Measuring Volume Resistance R
V
according to IEC 61340-4-1 /
DIN IEC 61340-4-1
(e.g. for measuring the volume resistance of materials)
➭
Connect the red measuring cable from the “+ pole” to the measuring
probe and the black, shielded measuring cable with the double
“COM / SHIELD” connection to the counter-electrode.
➭
Select the R
ESD
menu and a test voltage of U
M
= 10 volts. If a value of
less than 100 k
W
is displayed, the measured value is the result.
➭
If the value is displayed with R
X
> 100 k
W
, the test voltage must be
changed to 100 volts. The value then displayed is the result.
Test Voltage according to IEC 61340-5-1 / DIN EN 61340-5-1:
5.6
Measurement with the GUARD Terminal
The measurement of high-value resistance necessitates extremely small
measuring current and may be rendered problematic by influences such
as electromagnetic fields, humidity or surface current. A precise test setup
is therefore absolutely essential.
For measurements within a range of 10 G
W
to 1 T
W
, a guard cable must
be used to prevent surface current from distorting measurement results.
The GUARD terminal at the measuring probe prevents current at the sur-
face of a highly resistive test object from flowing from the + measurement
cable to the - measurement cable, instead of through the highly resistive
material itself.
Ring Probe in compliance with IEC 61340-5-1 / DIN EN 61340-5-1
➭
Connect the GUARD terminal at the measuring probe to the GUARD
jack at the METRISO
â
2000
for the measurement of highly
resistive materials.
Measurement is otherwise performed as described in the previous
chapters.
R
x
in
W
Test Voltage
(open-circuit voltage)
R
X
£
10
5
£
10 V
10
5
< R
X
£
10
12
100 V
R
x
in
W
Test Voltage
(open-circuit voltage)
R
X
£
10
5
£
10 V
10
5
< R
X
£
10
12
100 V
+ Pole
COM / SHIELD
Test Object
+
COM / SHIELD
Counter-Electrode
+ Pole
GUARD
COM / SHIELD