background image

Calibration Unit PGT120 

Part No.: 7100.PGT120.CU 

4 / 8 

2015-05-05 

3

 Footwear test verification (single shoe) - right 

To check the limits of the footwear test connect: 

-  the middle jack of the calibration unit with the 3 mm snap of the PGT

®

 120 

(same symbols 

 ). Use the DK3-socket adaptor which is included. 

-  the  left  jack  of  the  calibration  unit  with  the  red  4  mm  socket  on  the  rear  side  of  the 

PGT

®

 120 

(footwear electrode, same symbols    ). 

Set the marked lever of the rotary switch in succession to the positions mentioned below. 
Press the right contact electrode for each measurement. 

100k Lo 

100k OK 

35M OK 

35M Hi 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Display LED 

3

.1 

DIP-switch settings for upper limit 70 M

 

Switch 3 

Switch 4 

ON 

ON 

OFF 

OFF 

Set the marked lever of the rotary switch in succession to the positions mentioned below. 
Press the right contact electrode for each measurement. 

70M OK 

70M Hi 

 

 

 

 

 

 

 

 

 

 

 

 

Display LED 

Содержание 7100.PGT120.CU

Страница 1: ...Calibration Unit for PGT 120 Calibration manual for PGT120 with DIP12 switch...

Страница 2: ...4 OFF 8 3 2 1 2 Wrist strap verification To check the limits of the wrist strap test connect the middle jack of the calibration unit with the 3 mm snap of the PGT 120 same symbols Use the DK3 socket a...

Страница 3: ...h the red 4 mm socket on the rear side of the PGT 120 footwear electrode same symbols Set the marked lever of the rotary switch in succession to the positions mentioned below Press the right contact e...

Страница 4: ...the PGT 120 footwear electrode same symbols 4 1 DIP switch setting for upper limit 35 M Switch 3 Switch 4 ON 5 4 OFF ON OFF 8 3 2 1 Set the marked lever of the rotary switch in succession to the posit...

Страница 5: ...upper limit 140 M for series Switch 8 ON 8 5 3 ON OFF 4 2 1 Set the marked lever of the rotary switch in succession to the positions mentioned below Reset the instrument after each measurement by dis...

Страница 6: ...e Calibration Unit and set the marked lever of the rotary switch in succession to the marked positions The corresponding resistor values and tolerances can be taken from the drawing below Also connect...

Страница 7: ...S directive We identify our electrical and electronic devices as of August 2005 in accordance with WEEE 2012 19 EU and ElektroG with the symbol shown to the right per DIN EN 50419 These devices may no...

Отзывы: