background image

Calibration Unit PGT120 

Part No.: 7100.PGT120.CU 

3 / 8 

2015-05-05 

1

 Preparation 

Starting point for all measurements are the following DIP switch settings. 
The calibration is performed with the customers test voltage setting (DIP-switch 4 and 5) 
If desired, the test can be repeated with the other available test voltage settings. 

ON 

OFF 

2

 Wrist strap verification 

To check the limits of the wrist strap test connect: 
-  the middle jack of the calibration unit with the 3 mm snap of the PGT

®

 120 

 

(same symbols  

 ). Use the DK3-socket adaptor which is included. 

-  the left jack of the calibration unit with the black  4 mm banana socket of the PGT

®

 120. 

(wrist strap test, same symbols  

 ). 

Set the marked lever of the rotary switch in succession to the positions mentioned below. Press the 
left contact electrode for each measurement. 

750k Lo

 

750k OK

 

35M OK

 

35M Hi

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Display LED 

Содержание 7100.PGT120.CU

Страница 1: ...Calibration Unit for PGT 120 Calibration manual for PGT120 with DIP12 switch...

Страница 2: ...4 OFF 8 3 2 1 2 Wrist strap verification To check the limits of the wrist strap test connect the middle jack of the calibration unit with the 3 mm snap of the PGT 120 same symbols Use the DK3 socket a...

Страница 3: ...h the red 4 mm socket on the rear side of the PGT 120 footwear electrode same symbols Set the marked lever of the rotary switch in succession to the positions mentioned below Press the right contact e...

Страница 4: ...the PGT 120 footwear electrode same symbols 4 1 DIP switch setting for upper limit 35 M Switch 3 Switch 4 ON 5 4 OFF ON OFF 8 3 2 1 Set the marked lever of the rotary switch in succession to the posit...

Страница 5: ...upper limit 140 M for series Switch 8 ON 8 5 3 ON OFF 4 2 1 Set the marked lever of the rotary switch in succession to the positions mentioned below Reset the instrument after each measurement by dis...

Страница 6: ...e Calibration Unit and set the marked lever of the rotary switch in succession to the marked positions The corresponding resistor values and tolerances can be taken from the drawing below Also connect...

Страница 7: ...S directive We identify our electrical and electronic devices as of August 2005 in accordance with WEEE 2012 19 EU and ElektroG with the symbol shown to the right per DIN EN 50419 These devices may no...

Отзывы: