1.2 Electrical connection:

Insert the power cord into the socket 

(11)

 on the back panel of the unit, or into the socket of

the foot pedal. The default supply must be connected to a voltage source of 100 - 240 V /
50/60 Hz.

2. Operation: Flame regulation

The flame can be varied in size and intensity by turning the gas knob (

2

) and adjusting the

air knob (

2a

) to suit all requirements.

Attention:

 When operating the unit for the first time or after changing the nozzle, turn the gas

adjustment knob (

2

) two revolutions to the left and turn the air adjustment knob (

2a

) 3-4

revolutions to the left, too.

2.1 Operation: On-Off switch, operating mode

Switch the unit on by a short push on the function knob 

(1)

. It can be turned off by a long push

( 2 s) on the function knob.  By  turning the function knob all programs can be selected
and the IR sensor detection range can be adjusted.

2.2 Operation: Application programs

- BUTTON StartStop:

(LED Button is on)

The flame is ignited by operation of the function knob 

(1)

. The flame is extinguished after

renewed actuation of the function knob 

(1)

.Additionally the flame is automatically

extinguished when the burning timer has expired after 60 min.

- PEDAL Standard:

(LED Standard is on)

The flame is ignited by operation of the foot pedal. The foot pedal remains depressed for the
duration of use. The flame is extinguished upon release of the pedal.

- PEDAL Start-Stop:

(LED StartStop is on)

The flame is ignited by operation of the foot pedal. The flame is extinguished after renewed
actuation of the foot pedal. Additionally the flame is automatically extinguished when the
burning timer has expired after 60 min. Alternatively the flame can be extinguished by a short
push on the function knob 

(1)

.

Содержание gasprofi 1 SCS micro

Страница 1: ...The smallest laboratory gas burner with more safety EN Instruction manual DIN30665 part1 gasprofi 1 SCS micro...

Страница 2: ...Connector for foot pedal 11 Power connector for 9 V DC 12 Gas inlet R 1 4 L 13 Holding device for inoculation loop holder 14 Burner head screw 15 Monitor electrode 16 Ignition electrode 17 Active noz...

Страница 3: ...with two wrenches Ensure gas proofness with a suitable test fluid equipment DO NOT seal the swivel nut with Teflon tape etc Keep hands or other parts of the body away from the burner orifice 9a Do no...

Страница 4: ...lockwise Remove the active Nozzle N for natural gas 17 in the same way and exchange the nozzles ATTENTION Before mounting a nozzle check the O Ring 26 Replace the sealing if damaged or worn Art No 8 0...

Страница 5: ...By turning the function knob all programs can be selected and the IR sensor detection range can be adjusted 2 2Operation Applicationprograms BUTTON StartStop LED Button is on The flame is ignited by...

Страница 6: ...ubleClick IR Sensor 6 by passing you hand over it twice Whentheneededburningtimeisreachedpressthefunctionknob 1 tostoptheflame Thereby the reached burning time is memorized and the application program...

Страница 7: ...or programs in the button function and in the foot pedal operating mode StartStop is limited to 30 seconds see paragraph 2 2 If burning times longer than 30 seconds are required in case of a clogged b...

Страница 8: ...itor the burner head needs to be reinstalled prior a reset is possible 5 Cleaning and sterilizing Allow sufficient time for burner orifice 9 9a to cool down before disassembling or cleaning the burner...

Страница 9: ...he air knob 2a any longer During the use of the turbo flame most of the needed air is taken inside through the open burner shaft Remounting the cover of burner shaft see paragraph 5 2 7 Tilt adjustmen...

Страница 10: ...Notes...

Страница 11: ...rrectly Clean this areas with a brush The burner head can be cleaned with customary commercial disinfectants or it can be sterilized in an autoclave or washed in a dishwasher In operating mode Pedal S...

Страница 12: ...e see paragraph 4 Green LED Button Standard or StartStop blinks 3x Flame faliure see paragraph 4 Green LED Button Standard or StartStop blinks 4x Overtemperature see paragraph 4 Green LED Button Stand...

Страница 13: ...connection 1 ElektromagnetischeVertr glichkeit Electromagnetic Compatibility Directive 1 1 EN61326 1 2006 ElektrischeBetriebsmittelf rLeittechnikundLaboreinsatz EMV Anforderungen EN 61326 1 2006 Elect...

Страница 14: ...g h liquid gas Continuous cartridge operation CV 360 40 min Express 444 50 min CG 1750 150 min C 206 170 min CP 250 210 min CV 470 370min Temperatures Flame temperature 1350 C on liquid gas 1300 C on...

Страница 15: ...85 G ttingen Phone 49 0 551 793789 Fax 49 0 551 793707 Internet http www wld tec com 05 06 ProductionundService Halle KasselerStra e49 D 37318Arenshausen Phone 49 0 36081 68940 Fax 49 0 36081 68942 Em...

Отзывы: