WLD-TEC gasprofi 1 SCS micro Скачать руководство пользователя страница 14

Technical data:

Technology

Microprocessor

Programs

Foot pedal:

Standard (flame during pressed foot pedal)
Start-Stop with timer, 60 min

IR Sensor:

Start-Stop with timer, 60 min
Auto-Off with learn function, 1 sec - 60 min

Button:

Start-Stop with timer, 60 min

Safety features

Safety Control System (SCS):

ignition, flame and temperature monitor

with gas safety cut off

burner  head  clogging  and
assembly monitor (BHC)
automatic unit switch off, 4h
residual heat display

Gas supply and consumtion

Gas supply:

1/4“ left + filter

Gas types:

II2ELL3B/P: natual gas E/LL,18 - 25 mbar
liquid gas, 20 - 50 mbar

Connected load:

70 g/h liquid gas

Continuous  cartridge  operation:

CV 360 - 40 min,

Express 444 - 50 min,

CG 1750 - 150 min, C 206 - 170 min,
CP 250 - 210 min,

CV 470 - 370min

Temperatures

Flame temperature:

1350 °C on liquid gas
1300 °C on natural gas (E)

Temperature  threshold  level:

1 kW liquid gas, 1 kW natural gas

Electrical

Power consumption:

2 VA

Power  connection:

100 - 240 V / 50/60 Hz / max. 0.3 A
9 V DC / 1 A

IR-Sensor  coverage:

35 - 70 mm, adjustable

DoubleClick  IR-Sensor:

Time range 1 sec

Mechanical

Casing and operating controls:

stainless steel / glass, UV and solvent resistant

Burner head:

removable and decomposable, stainless steel

Cover of the burner shaft:

Ø 23 mm, with drains

Measurements (B x H x T):

85 x 49 x 86 mm

Weight:

700 g

Licences

DIN-DVGW Reg.-Nr.:

NG-2211AS0167

CE:

EN 61326-1, EN 61010-1

EEC guidelines:

2004/108/EC and 73/23/EEC

Содержание gasprofi 1 SCS micro

Страница 1: ...The smallest laboratory gas burner with more safety EN Instruction manual DIN30665 part1 gasprofi 1 SCS micro...

Страница 2: ...Connector for foot pedal 11 Power connector for 9 V DC 12 Gas inlet R 1 4 L 13 Holding device for inoculation loop holder 14 Burner head screw 15 Monitor electrode 16 Ignition electrode 17 Active noz...

Страница 3: ...with two wrenches Ensure gas proofness with a suitable test fluid equipment DO NOT seal the swivel nut with Teflon tape etc Keep hands or other parts of the body away from the burner orifice 9a Do no...

Страница 4: ...lockwise Remove the active Nozzle N for natural gas 17 in the same way and exchange the nozzles ATTENTION Before mounting a nozzle check the O Ring 26 Replace the sealing if damaged or worn Art No 8 0...

Страница 5: ...By turning the function knob all programs can be selected and the IR sensor detection range can be adjusted 2 2Operation Applicationprograms BUTTON StartStop LED Button is on The flame is ignited by...

Страница 6: ...ubleClick IR Sensor 6 by passing you hand over it twice Whentheneededburningtimeisreachedpressthefunctionknob 1 tostoptheflame Thereby the reached burning time is memorized and the application program...

Страница 7: ...or programs in the button function and in the foot pedal operating mode StartStop is limited to 30 seconds see paragraph 2 2 If burning times longer than 30 seconds are required in case of a clogged b...

Страница 8: ...itor the burner head needs to be reinstalled prior a reset is possible 5 Cleaning and sterilizing Allow sufficient time for burner orifice 9 9a to cool down before disassembling or cleaning the burner...

Страница 9: ...he air knob 2a any longer During the use of the turbo flame most of the needed air is taken inside through the open burner shaft Remounting the cover of burner shaft see paragraph 5 2 7 Tilt adjustmen...

Страница 10: ...Notes...

Страница 11: ...rrectly Clean this areas with a brush The burner head can be cleaned with customary commercial disinfectants or it can be sterilized in an autoclave or washed in a dishwasher In operating mode Pedal S...

Страница 12: ...e see paragraph 4 Green LED Button Standard or StartStop blinks 3x Flame faliure see paragraph 4 Green LED Button Standard or StartStop blinks 4x Overtemperature see paragraph 4 Green LED Button Stand...

Страница 13: ...connection 1 ElektromagnetischeVertr glichkeit Electromagnetic Compatibility Directive 1 1 EN61326 1 2006 ElektrischeBetriebsmittelf rLeittechnikundLaboreinsatz EMV Anforderungen EN 61326 1 2006 Elect...

Страница 14: ...g h liquid gas Continuous cartridge operation CV 360 40 min Express 444 50 min CG 1750 150 min C 206 170 min CP 250 210 min CV 470 370min Temperatures Flame temperature 1350 C on liquid gas 1300 C on...

Страница 15: ...85 G ttingen Phone 49 0 551 793789 Fax 49 0 551 793707 Internet http www wld tec com 05 06 ProductionundService Halle KasselerStra e49 D 37318Arenshausen Phone 49 0 36081 68940 Fax 49 0 36081 68942 Em...

Отзывы: