Section 12 : Technical Support :
My Sensor Isn't Reading Correctly
If your sensor is not reading the correct values, follow these steps.
1. Make sure the sensor is suitable for your system. Refer to the
"Specifications" section (pages 8 — 10 for details)
2. Make sure the sensor is calibrated. Sensors should be calibrated
every 2 years. Contact your local dealer or WiseAir Technologies for calibration.
3. Make sure the sensor is installed correctly. Refer to "Installation —
Mechanical" section (pages 11 - 14 for more details)
The sensor should be installed at 90° to the pipe.
Typical installation Errors include:
Installing sensor upside down or at an angle
Installing sensor in wet air
Air is not reaching the sensor tip (sensor tip is not far enough into pipe).
4. Make sure the sensor is wired correctly. Refer to “Installation —
Electrical (Wiring)” section (page 15) for more information.
5. Check Modbus Settings. Refer to “Communication Settings “section
(page 16) for more information.
6. Check dryer is functioning correctly
7. Check condensate drains are functioning correctly
8. Is your associated equipment compatible with the dew point sensor?
If you are still having problems, contact your local dealer or WiseAir
Technologies.