WINTACT WT100A Скачать руководство пользователя страница 8

13

Avoid measurement error

11

12

12.5. Measurement of composite shape:
    When measuring materials of composite shapes (such as 
elbow of pipe), the method described in 12-4 can be adopted, 
except that the measurement should be performed for two 
times to get two readings by placing the crosstalk spacer of 
probe vertical or parallel with the axis, the smaller of which is 
the thickness of the measured point.

12.6. Non-parallel surfaces:
    In order to obtain a satisfactory ultrasonic response, the other 
surface of the material to be tested must be parallel or coaxial 
with the surface to be measured, otherwise it will cause 
measurement errors or no reading.

12.7. Effects of material's temperature:
    The thickness of material and the transmission speed of 
ultrasonic wave are affected by temperature. For high 
measurement accuracy, separately measure samples of the 
same material under the same temperature condition, then 
calculate the measurement error caused by temperature, and 
correct error with parameterS. For steel, high temperatures will 
cause big error (measured values are smaller than actual data), 
and this way can also be used to correct errors.

12.8. Materials of attenuation:
    For some materials with fibers, pores, and coarse particles, 
they cause great scattering and energy attenuation of 
ultrasonic waves, resulting in abnormal readings or even no 
readings (usually the abnormal reading is smaller than the 
actual thickness), in which case the material is not suitable to 
be tested with this thickness gauge.

12.9. Samples for reference:
    A material of known thickness or sound speed helps calibrate 
the instrument. At least one sample for reference is required to 
calibrate the instrument. The ultrasonic thickness gauge is 
equipped with a steel sample with thickness of 4.0 mm. Please 
refer to calibration instruction in item 9 of this manual for 
calibration method.

13.1. Ultra-thin material
    With any ultrasonic thickness gauge, if the thickness of the 
tested material falls below the probe's lower limit, measurement 
errors will appear. If necessary, the minimum thickness limit can 
be measured by sample comparison.
    When measuring ultra-thin materials, sometimes an error 
called "double refraction" occurs, in which the measured value is 
equal to twice the actual thickness. Another error is called "pulse 
envelope and loop jump", in which the measured value is bigger 
than the actual thickness. To avoid such errors, repeat checking 
measurement of critical thin material.

13.2. Rust spots and corrosion pits
    Rust spots and corrosion pits on the other surface of the tested 
material will cause readings to change irregularly or no reading 
in extreme cases, and small rust spots are sometimes difficult to 
detect. When pits are found or suspected, be careful with the 
measurement of this area, and place the crosstalk spacer of 
probe at different angles for multiple tests.

13.3. Material identification error
    If you have calibrated the instrument with one material and use it 
to test another material, an erroneous result will occur. You 
should be careful when selecting the correct sound speed. Or, if 
the sound speed in the actual test displays certain deviation from 
that of calibration, errors may also occur. Please make 
adjustment in use.

13.4. Probe abrasion
    The surface of probe is made of acrylic resin. After the 
instrument is used for a long time, the roughness of probe 
surface will increase, resulting in a decrease of sensitivity. If the 
user can determine errors caused by this, the probe surface can 
be polished with a small amount of 500# sandpaper or whetstone 
to smooth surface and ensure parallelism.
If the result is still unstable, you may need to replace the probe.

Содержание WT100A

Страница 1: ...eserves the right to modify product design and specification without notice Version WT100A WT130A EN 02 MODEL WT100A WT130A Ultrasonic Thickness Gauge MENU OK VEL UP ESC DOWN Ultrasonic Thickness Gaug...

Страница 2: ...nctions 06 Measurement options 06 Probe 06 Limit value 06 Storage 06 Chart 07 Settings 07 12 10 Technology of measurement Menu interface display instruction 03 8 07 Bottom menu operation and function...

Страница 3: ...tion and other testing areas Any material that allows ultrasonic to transmit at a constant speed and that is distinguishable from waves reflected from the back side is applicable to this instrument 1...

Страница 4: ...be Measurement parameters 1 0 300 0mm 1 0 50 0mm 1 2 300 0mm 1 2 300 0mm 1 0 225 0mm WT130A WT100A Measurement parameters Measuring range Data storage Probe selection WT130A 1 00 300 0mm 1500 WT100A 1...

Страница 5: ...ment settings under this menu including a Sound speed of measurement sound speed bar of measurement interface b Measurement unit of c Measurement resolution 2 Probe Under this menu select the probe to...

Страница 6: ...to bottom menu functions a Storage Save current measured values to the selected file b Delete Delete measured values and measurement information except for limit values c Calibration Refer to calibra...

Страница 7: ...be as smooth as possible through sanding grinding polishing rasping or using high viscosity coupling agent 12 3 Surface made by crude machinery Regular fine grooves on surfaces made by crude machiner...

Страница 8: ...h a steel sample with thickness of 4 0 mm Please refer to calibration instruction in item 9 of this manual for calibration method 13 1 Ultra thin material With any ultrasonic thickness gauge if the th...

Страница 9: ...omly mixed coupling agent light machine oil etc are suitable For rough surface vertical surface and top surface or aluminum higher viscosity coupling agents such as glycerin creams greases etc can be...

Страница 10: ...sides the maintenance fee will be charged according to our company s regulations The maintenance fee for special configuration out of our company s regular products will be charged according to releva...

Отзывы: