8.1.

 

Introduction to bottom menu operation;

a

Enter into bottom menu:

   

Press ESC under measurement 

      interface to enter into bottom 
      menu.

  

b

Select menu item:

   

Press UP or DOWN button to 

      select menu item.
c

Enter into selected menu item:

   

Press OK to enter into selected 

     menu item.
d

Exit menu item:

   

Press ESC button to exit bottom menu.

5

Chart

  

In this menu, you can view the data saved in the file inform of 

trend graph.

   

a. View basic information such as data number, maximum 

          value, minimum value, and average value in this file.

   

b. Zoom the whole trend graph

   

c. Use cursor to view specific data value in the file.

6

Settings

  

Make other settings for the instrument under this menu, including:

   

a. Language 

     b. Automatic shutdown time  

     c. Button sound  

     d. Backlight brightness  

     e. Color  

     f.  Software version number 

     g. Restore factory settings

Bottom menu

8

Bottom menu operation and 

 

function introduction

8.2. Introduction to bottom menu functions;

  

a. Storage:     

       Save current measured values to the selected file.
   b. Delete:    
       Delete measured values and measurement information 
       (except for limit values).
   c. Calibration:    
       Refer to calibration instruction for details.
   d. Sound speed calibration:    
       Refer to sound speed calibration instruction for details.

9.1. Press ESC under measurement interface 
        to enter into bottom menu. Select 
        calibration item and press OK button.
9.2. The interface is displayed as shown on 
        the right:
9.3. The sound speed in calibration is 
        automatically set as 5900m/s during.
9.4. Press the probe on calibration sample to 
        calibrate. 5 circles on the bottom of 
        screen will be filled in turn.
9.5. After the calibration is completed, it will 
        automatically return to measurement 
        interface.

10.1. Enter into bottom menu and select 
          sound speed calibration item 
10.2. The interface is displayed as shown 
          on the left:
10.3. As shown in the figure, press 
         UP/DOWN  button to adjust value.
10.4. After the value adjustment is 
         completed, press OK to complete 
         calibration.

9

Calibration instruction

10

Instruction of sound speed calibration

Содержание WT100A

Страница 1: ...eserves the right to modify product design and specification without notice Version WT100A WT130A EN 02 MODEL WT100A WT130A Ultrasonic Thickness Gauge MENU OK VEL UP ESC DOWN Ultrasonic Thickness Gaug...

Страница 2: ...nctions 06 Measurement options 06 Probe 06 Limit value 06 Storage 06 Chart 07 Settings 07 12 10 Technology of measurement Menu interface display instruction 03 8 07 Bottom menu operation and function...

Страница 3: ...tion and other testing areas Any material that allows ultrasonic to transmit at a constant speed and that is distinguishable from waves reflected from the back side is applicable to this instrument 1...

Страница 4: ...be Measurement parameters 1 0 300 0mm 1 0 50 0mm 1 2 300 0mm 1 2 300 0mm 1 0 225 0mm WT130A WT100A Measurement parameters Measuring range Data storage Probe selection WT130A 1 00 300 0mm 1500 WT100A 1...

Страница 5: ...ment settings under this menu including a Sound speed of measurement sound speed bar of measurement interface b Measurement unit of c Measurement resolution 2 Probe Under this menu select the probe to...

Страница 6: ...to bottom menu functions a Storage Save current measured values to the selected file b Delete Delete measured values and measurement information except for limit values c Calibration Refer to calibra...

Страница 7: ...be as smooth as possible through sanding grinding polishing rasping or using high viscosity coupling agent 12 3 Surface made by crude machinery Regular fine grooves on surfaces made by crude machiner...

Страница 8: ...h a steel sample with thickness of 4 0 mm Please refer to calibration instruction in item 9 of this manual for calibration method 13 1 Ultra thin material With any ultrasonic thickness gauge if the th...

Страница 9: ...omly mixed coupling agent light machine oil etc are suitable For rough surface vertical surface and top surface or aluminum higher viscosity coupling agents such as glycerin creams greases etc can be...

Страница 10: ...sides the maintenance fee will be charged according to our company s regulations The maintenance fee for special configuration out of our company s regular products will be charged according to releva...

Отзывы: