![Winncare Medicalys MS4/C Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/winncare/medicalys-ms4-c/medicalys-ms4-c_manual_987325012.webp)
12
6. USE
6.1. Purpose of bed
➢
Depending on their configuration, Medicalys beds are intended for environments 1, 2, 3 and 5 defined
by the standard. EN 60601-2-52.
➢
Medicalys are intended for persons whose size is greater than or equal to 146 cm, with a weight
greater than or equal to 40 kg and a body mass index (BMI) greater than or equal to 17.
6.2. Specific precautions for use
The bed can be used as a stretcher except with wheels Ø100.
6.2.1. Residual risks
6.3. General description
Manual crank or electric folding leg rest depending on version.
DANGER
RESULT
REQUIREMENTS FOR THE USER
Trapping Hands pinched
Handle legsrest or backrest up by handles or at specified
locations
Trapping Crushing body or object
Before lowering the bed, check that the area between
the bed and the floor is free
Trapping Crushing body or object
Check before operating the central brake pedal that the
area between the pedal and the floor is free
Slatted back rest
Mattress side stop
Bed base
Headboard
Angled lifting
pole slots
IV stand slots
Footboard
Wall stop
3 positions
Back rest handle
Base and
cross braces
Forward slant /
back slant pedal
Centralised
brake pedal
Slat leg rest
Leg rest handle
Calculation of BMI :
BMI = m / h²
m : weight in kg
h : size in m
Warning refer to
user’s manual