681XXB rear panel
IEEE-488 GPIB
con-
nector.
Step 3
Turn on the 681XXB and the 562. The
system is now ready to operate.
Using Discrete Cables
Connect the 681XXB to the 562 SNA as follows:
Step 1
Connect one end of a coaxial cable to the
681XXB rear panel
HORIZ OUT
connec-
tor. Connect the other end of the cable to
the 562 rear panel
HORIZONTAL INPUT
/OUTPUT
connector.
Step 2
Connect one end of a coaxial cable to the
681XXB rear panel
SEQ SYNC OUT
con-
nector. Connect the other end of the cable
to the 562 rear panel
SEQ SYNC INPUT
connector.
Step 3
Connect one end of a coaxial cable to the
681XXB rear panel
MARKER OUT
connec-
tor. Connect the other end of the cable to
the 562 rear panel
VIDEO MARKER IN-
PUT
connector.
Step 4
Connect one end of a coaxial cable to the
681XXB rear panel
DWELL IN
connector.
Connect the other end of the cable to the
562 rear panel
SWEEP DWELL OUTPUT
connector.
Step 5
Connect one end of the dedicated system
bus cable (P/N 2100-1) to the 562 rear
panel
DEDICATED GPIB
connector. Con-
nect the other end of the cable to the
681XXB rear panel
IEEE-488 GPIB
con-
nector.
Step 6
Turn on the 681XXB and the 562. The
system is now ready to operate.
USE WITH
OPERATION WITH A 562
OTHER INSTRUMENTS
SCALAR NETWORK ANALYZER
NOTES
The 562 SNA will only accept and
display the nine video markers, F1
thru F9, from the 681XXB.
When performing amplifier testing
only use the 681XXB power level,
L1.
681XXB OM