6-4
ROUTINE MAINTENANCE
Routine maintenance that can be performed by the operator consists of
replacing a defective line fuse, cleaning the fan filter, and cleaning the
data display.
Replacing
the Line Fuse
The value of the line fuse used in the 681XXB is de-
termined by the line voltage selection—a 5A line
fuse for 110 Vac line voltage; a 2.5A line fuse for
220 Vac line voltage. These line fuse values are
printed on the rear panel next to the fuse holder.
Step 1
Disconnect the 681XXB from the power
source.
Step 2
Using a small flat-blade screwdriver, turn
the fuse cap counter-clockwise and re-
move the fuse holder.
Step 3
Replace the fuse in the fuse holder.
Step 4
Install the fuse holder in the rear panel.
Using the screwdriver, rotate the fuse cap
clockwise to secure the fuse holder in
place.
Step 5
Reconnect the sweep generator to the
power source.
OPERATOR
ROUTINE
MAINTENANCE
MAINTENANCE
Fuse
Fuse Holder
Figure 6-1.
Replacing the Line Fuse
681XXB OM
6-13