Willow Breast Pump Instructions for Use
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSIONS
The Willow Breast Pump is intended for use in the electromagnetic environment specified below.
The customer or user of the Breast Pump should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic Environment –
Guidance
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
<5 % UT (>95 % dip
in UT) for 0.5 cycle
30% reduction,
25/30 periods at 0°
> 95% reduction,
0.5 period at 0°,
45°, 90°, 135°, 180°,
225°, 270° and 315°
> 95% reduction, 1
period at 0°
> 95% reduction,
250/300 periods
<5 % UT (>95 % dip
in UT) for 0.5 cycle
30% reduction,
25/30 periods at 0°
> 95% reduction,
0.5 period at 0°,
45°, 90°, 135°, 180°,
225°, 270° and 315°
> 95% reduction, 1
period at 0°
> 95% reduction,
250/300 periods
The Willow Breast Pump is suitable
for use in all establishments,
including domestic establishments
and those directly connected to the
public low-voltage power supply
network that supplies buildings used
for domestic purposes.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
The Willow Breast Pump is suitable
for use in all establishments,
including domestic establishments
and those directly connected to the
public low-voltage power supply
network that supplies buildings used
for domestic purposes.
NOTE UT is the a.c. mains voltage prior to application of the test level.
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSIONS
The Willow Breast Pump is intended for use in the electromagnetic environment specified below.
The customer or user of the Breast Pump should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment –
Guidance
Conducted RF
IEC 61000-4-6
3 V; 150 kHz
to 80 MHz
6 V; ISM and
amateur
radio bands
between
0.15 MHz and
80 MHz
3 V; 150 kHz
to 80 MHz
6 V; ISM and
amateur
radio bands
between
0.15 MHz and
80 MHz
Portable and mobile RF
Communications equipment should be used
no closer to any part of the Willow Breast Pump,
including cables, than the recommended separation
distance calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
d = 1.2
√
P
d = 1.2
√
P 80 MHz to 800 MHz
d = 2.3
√
P 800 MHz to 2.5 GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a.
should be less than the compliance level in each
frequency range.
b.
Interference may
occur in the vicinity of equipment
marked with the following symbol:
Radiated RF IEC
61000-4-3
3V/m
80 MHz to
2.7 GHz
3V/m
80 MHz to
2.7 GHz
NOTE 1: At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.