GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSIONS

The Willow Breast Pump is intended for use in the electromagnetic environment specified below.  
The customer or user of the Breast Pump should assure that it is used in such an environment.

Immunity Test

IEC 60601  
Test Level

Compliance Level

Electromagnetic Environment –
Guidance

Voltage 
dips, short 
interruptions 
and voltage 
variations on 
power supply 
input lines
IEC 61000-4-11

<5 % UT (>95 % dip 
in UT) for 0.5 cycle
30% reduction, 
25/30 periods at 0°
> 95% reduction, 
0.5 period at 0°, 
45°, 90°, 135°, 180°, 
225°, 270° and 315° 
> 95% reduction, 1 
period at 0°
> 95% reduction, 
250/300 periods

<5 % UT (>95 % dip 
in UT) for 0.5 cycle
30% reduction, 
25/30 periods at 0°
> 95% reduction, 
0.5 period at 0°, 
45°, 90°, 135°, 180°, 
225°, 270° and 315° 
> 95% reduction, 1 
period at 0°
> 95% reduction, 
250/300 periods

The Willow Breast Pump is suitable 
for use in all establishments, 
including domestic establishments 
and those directly connected to the 
public low-voltage power supply 
network that supplies buildings used 
for domestic purposes.

Power 
frequency 
(50/60 Hz) 
magnetic field
IEC 61000-4-8

30 A/m

30 A/m

The Willow Breast Pump is suitable 
for use in all establishments, 
including domestic establishments 
and those directly connected to the 
public low-voltage power supply 
network that supplies buildings used 
for domestic purposes.

NOTE UT is the a.c. mains voltage prior to application of the test level.

GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSIONS

The Willow Breast Pump is intended for use in the electromagnetic environment specified below.  
The customer or user of the Breast Pump should assure that it is used in such an environment.

Immunity Test

IEC 60601 
Test Level

Compliance 
Level

Electromagnetic Environment –
Guidance

Conducted RF
IEC 61000-4-6

3 V; 150 kHz
to 80 MHz

6 V; ISM and 
amateur
radio bands 
between
0.15 MHz and 
80 MHz

3 V; 150 kHz
to 80 MHz

6 V; ISM and 
amateur
radio bands 
between
0.15 MHz and 
80 MHz

Portable and mobile RF
Communications equipment should be used 
no closer to any part of the Willow Breast Pump, 
including cables, than the recommended separation 
distance calculated from the equation applicable to 
the frequency of the transmitter.
Recommended separation distance
d = 1.2

P

d = 1.2

P 80 MHz to 800 MHz  

d = 2.3

P 800 MHz to 2.5 GHz

Where P is the maximum output power rating of the 
transmitter in watts (W) according to the transmitter 
manufacturer and d is the recommended 
separation distance in meters (m).
Field strengths from fixed RF transmitters, as 
determined by an electromagnetic site survey,  

a.

 should be less than the compliance level in each 

frequency range. 

b.

 Interference may  

occur in the vicinity of equipment  
marked with the following symbol:

Radiated RF IEC 
61000-4-3

3V/m 
80 MHz to 
2.7 GHz

3V/m 
80 MHz to 
2.7 GHz 

NOTE 1: At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by 
absorption and reflection from structures, objects and people.

Содержание Willow Pump

Страница 1: ...Instructions for Use...

Страница 2: ......

Страница 3: ...IFU 1004 J...

Страница 4: ...ange Flextube 24 Wake Up Willow 24 Assembly 25 B Pumping 30 Turn on Pump 30 Alignment 31 Initiate Latch 34 Secure Pump 35 Pause Pump 36 Break Latch Remove Pump 36 Flip to Finish 38 Turn Pump Off 42 Op...

Страница 5: ...ect Pump Flange and Flextube to ensure there are no tears or indications of degraded performance 4 If excessive discomfort or pain is felt during pumping stop pumping break latch then reposition Pump...

Страница 6: ...leaned prior to use reference Section 7A GET READY Boil Flange Flextube These parts are intended for use by a single user only and should be replaced every 3 months Milk Bags are intended for single u...

Страница 7: ...Instructions for Use ALIGNMENT The goal of Alignment is to center your nipple in the Flange Tunnel so that it doesn t touch any sides of the Flange Make sure the edges are flush Click No smooshed nipp...

Страница 8: ...t our website www willowpump com Your breast flange size may fluctuate with birth breastfeeding and pumping We recommend you revisit your flange fit from time to time Too Small During pumping some or...

Страница 9: ...Keys to success Pressing Pump against breast while initiating latch Picking the right bra While in Stimulation lean back until you reach Expression YOUR BRA The bra you wear while pumping with Willow...

Страница 10: ...quiet sounds Many women transition within 5 7 minutes but this can vary Suction level starts at 1 but you can adjust to your desired level Note you cannot go back to Stimulation phase under a minute...

Страница 11: ...Use FLIP TO FINISH Willow s unique emptying feature is called Flip to Finish You will want to check it out before your first pumping session 6 PRODUCT PARTS Additional Flange sizes and spare parts ar...

Страница 12: ...Port Reset Button Suction Controls Increase Decrease On Off Flange Exterior Tunnel Alignment Guides Door Button on both sides secure Flange to Pump Flange Interior Button on both sides secure Flange t...

Страница 13: ...Y WARNINGS DO NOT wash or sanitize Pump Wipe with clean damp cloth if needed DO NOT use harsh chemicals to clean Flange and Flextube Use mild dish washing soap AFTER EVERY USE FOLLOW THESE DIRECTIONS...

Страница 14: ...ess avoid touching inside of parts fig 3d Insert small brush into Tip at bottom and clean fig 3c Use small brush to clean inside top opening and area under Collar IMPORTANT Take care not to damage par...

Страница 15: ...DO NOT microwave breast milk DO NOT use hot water to thaw frozen breast milk DO NOT place filled Milk Bag in refrigerator door to avoid temperature fluctuations CHARGING WILLOW 1 Plug Charger into Pu...

Страница 16: ...ump b Wait until Pump lights up c Unplug Charger from Pump fig 1a Hold Pump upright then open by squeezing the two buttons on the top sides of Flange fig 1b Pull top open first then unhook Tab and Slo...

Страница 17: ...or open by pushing the blue thumb pad in and sliding up b Attach Milk Bag fig 3e Drape a new Milk Bag over Flange with writing THIS SIDE UP facing away from Flange Firmly press Milk Bag s One way Valv...

Страница 18: ...wn fig 3h The Milk Bag is bigger than the Flange To fit it press center of bag into deepest part of Flange then tuck edges TIP Do not fold Milk Bag to make it fit fig 3i Tuck bag under tabs on Flange...

Страница 19: ...OFF Press On Off button until Pump lights up Color of light indicates battery status Green Solid Enough charge for 2 sessions Orange Solid Enough charge for 1 session Red Solid Low battery inadequate...

Страница 20: ...may not have the correct Flange size ONCE YOU START PUMPING YOU MAY NEED TO REALIGN IF YOU Experience excessive discomfort or pain that is not alleviated by pressing Willow firmly against your breast...

Страница 21: ...nd slower the Pump has detected let down and is collecting milk in the Milk Bag Suction level can be adjusted Suction level starts at 1 reference Section 7C SUCTION LEVELS IMPORTANT Always break latch...

Страница 22: ...Milk Bag When Milk Bag is full Pump stops pumping Break latch remove Pump perform Flip to Finish reference Section 7B PUMPING Flip to Finish and replace Milk Bag To resume pumping attach a new Milk Ba...

Страница 23: ...ilk Bag the Flange starts to separate from the Pump This is normal fig 7a With Flange side up ROTATE so blue thumb pad is pointing toward you TIP It is completely normal for a few drops to be left in...

Страница 24: ...n the Flange to open and slowly lift the Flange Be careful not to tip the Flange too far fig 7h If more than a few drops are present turn Pump back on press the Pump Pause button and repeat Flip to Fi...

Страница 25: ...d Flextube as instructed reference Section 6 PRODUCT PARTS CLEANING MAINTENANCE 12 CHARGE Check battery level after each use by pressing On Off button If the battery level is low charge Pump with Char...

Страница 26: ...to maximum level of suction that is still comfortable 8 TROUBLESHOOTING Use table to help solve any problems that may arise If problem persists or to report unexpected operation or events call Custome...

Страница 27: ...n break latch and remove Pump OTHER ISSUES TO RESOLVE ISSUE Pump does not pump Check that Pump is not plugged into Charger Confirm that Pump is powered on by pressing the On Off button until Pump ligh...

Страница 28: ...d replace if damaged 3 Check that power is on and battery has sufficient charge 4 Press Pump securely against breast Note Suction level can only be adjusted in Expression phase Pump stops pumping Pump...

Страница 29: ...und by Willow to not meet the above limited warranty in the Warranty Period as purchaser s sole and exclusive remedy Willow will repair or at the option of Willow replace this product without charge f...

Страница 30: ...medical procedure or during imaging For example do not use pump during RF ablation of a skin anomaly x ray MRI or ultrasound With other Equipment Use of this equipment adjacent to or stacked with othe...

Страница 31: ...there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception which can be determined by turni...

Страница 32: ...3 Complies GUIDANCE AND MANUFACTURER S DECLARATION ELECTROMAGNETIC EMISSIONS The Willow Breast Pump is intended for use in the electromagnetic environment specified below The customer or user of the...

Страница 33: ...he a c mains voltage prior to application of the test level GUIDANCE AND MANUFACTURER S DECLARATION ELECTROMAGNETIC EMISSIONS The Willow Breast Pump is intended for use in the electromagnetic environm...

Страница 34: ...observed additional measures may be necessary such as reorienting or relocating the Willow Breast Pump b Over the frequency range 150 kHz to 80MHz field strengths should be less than 3 V m 11 OPERATI...

Страница 35: ...n Catalog Number Atmospheric Limitation Serial Number Consult Instructions For Use Do not use if package is damaged Caution SYMBOLS SYMBOL DEFINITION SYMBOL DEFINITION Keep dry Type BF Applied Part Cl...

Страница 36: ...reserved Willow Flextube and the stylized W are trademarks or registered trademarks of Willow in the United States and other countries Patent www willowpump com patents CAREFULLY READ ALL INSTRUCTIONS...

Отзывы: