Willow Breast Pump Instructions for Use
Warning:
Use of accessories other than those specified by Exploramed NC7 may result in
increased emissions or decreased immunity of the Willow Breast Pump.
The Willow Breast Pump has no essential performance; however, it was tested for immunity to
electromagnetic disturbances. The Willow Breast Pump passed with the following acceptance
criteria: pumping vacuum does not exceed a safe level and no components fail.
This equipment uses wireless communication:
Frequency (transmit and receive): 2400 – 2483.5 MHz
Modulation: frequency hopping with GFSK
Effective radiated power: 4 mW
This device complies with part 15 of the FCC Rules. Operation is subject to the following two
conditions: (1) This device may not cause harmful interference, and (2) this device must accept
any interference received, including interference that may cause undesired operation.
Note:
This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference in a residential installation. This equipment generates,
uses and can radiate radio frequency energy and, if not installed and used in accordance with
the instructions, may cause harmful interference to radio communications. However, there is
no guarantee that interference will not occur in a particular installation. If this equipment does
cause harmful interference to radio or television reception, which can be determined by turning
the equipment off and on, the user is encouraged to try to correct the interference by one or
more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different
from that to which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
ELECTROMAGNETIC COMPATIBILITY COMPLIANCE STATEMENT INFORMATION
Recommended separation distances between portable and mobile RF communications equipment
and the Willow Breast Pump
Rated maximum output
power of transmitter
Separation distance according to frequency of transmitter (m)
w
150kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.69
3.69
7.38
100
11.67
11.67
23.33
For transmitters rated at a maximum output power not listed above, the recommended separation distance d
in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.