InfraCal TOG/TPH Analyzer,
Model CVH
Calibration
For questions please contact us at:
[email protected], or 203-855-9136
Содержание InfraCal CVH
Страница 1: ...InfraCal TOG TPH Analyzer Model CVH Calibration www WilksIR com...
Страница 2: ...Calibration Oil Heavy hydrocarbon No volatiles No volatiles No Additives...
Страница 4: ...First Standard 100 ppm Add 10 ml of solvent to the 10 ml graduated cylinder...
Страница 7: ...Second Standard 50 ppm Pour 5 ml of the 100 ppm standard back into the 10 ml graduated cylinder...
Страница 13: ...Zeroing Fill the cuvette with clean solvent and place in the sample stage with the frosted side facing front...
Страница 14: ...Zeroing Press and hold the ZERO button until the display reads bAL Recall ZERO CAL RUN bAL...
Страница 16: ...Zeroing Press RUN The display should read 00 02 If not repeat the zero process Recall ZERO CAL RUN 00...
Страница 18: ...Example Calibration Table A Absorbance C Calibration Standard AO1 42 CO1 25 AO2 81 CO2 50 AO3 149 CO3 100...
Страница 20: ...Press the CAL button for two seconds until CAL appears on the display CAL Recall ZERO CAL RUN...
Страница 21: ...Press RECALL until Edit is displayed Edit Recall ZERO CAL RUN...
Страница 22: ...3 Press CAL and the display will read n for a short time n 3 Recall ZERO CAL RUN n...
Страница 23: ...3 Set n to the number standards using the up and down arrows n 3 Recall ZERO CAL RUN n...
Страница 25: ...Press CAL and the display will read A01 for a short time A01 Recall ZERO CAL RUN A01...
Страница 28: ...Press CAL again and the display will read C01 C01 Recall ZERO CAL RUN C01...
Страница 33: ...Do the same for AO3 and CO3 Absorbance Calibration Standard AO1 42 CO1 25 AO2 81 CO2 50 AO3 149 CO3 100...