background image

25

Using the up and down arrows, scroll to the first    

calibration value 

(for our example it will be 25)

Recall

ZERO

CAL

RUN

25

Содержание InfraCal CVH

Страница 1: ...InfraCal TOG TPH Analyzer Model CVH Calibration www WilksIR com...

Страница 2: ...Calibration Oil Heavy hydrocarbon No volatiles No volatiles No Additives...

Страница 3: ...Mixing Standards Volume volume ppm Oil Solvent 400 40ul 10 ml 300 30ul 10 ml 300 30ul 10 ml 200 20ul 10 ml 100 10 ul 10 ml...

Страница 4: ...First Standard 100 ppm Add 10 ml of solvent to the 10 ml graduated cylinder...

Страница 5: ...First Standard 100 ppm Add 10 ml of solvent to the 10 ml graduated cylinder Using the syringe add 10 microliters of oil to 10 ml of solvent in the ml of solvent in the graduated cylinder...

Страница 6: ...Add 10 ml of solvent to the 10 ml graduated cylinder Using the syringe add 10 microliters of oil to 10 ml of solvent in the ml of solvent in the graduated cylinder Pour this standard into a vial label...

Страница 7: ...Second Standard 50 ppm Pour 5 ml of the 100 ppm standard back into the 10 ml graduated cylinder...

Страница 8: ...Second Standard 50 ppm Pour 5 ml of the 100 ppm standard back into the 10 ml graduated cylinder Fill the cylinder to 10 ml with clean solvent...

Страница 9: ...5 ml of the 100 ppm standard back into the 10 ml graduated cylinder Fill the cylinder to 10 ml with clean solvent This will cut the 100 ppm standard in half to be 50 ppm Pour this standard into a vial...

Страница 10: ...Third Standard 25 ppm Pour 5 ml of the 50 ppm standard back into the 10 ml graduated cylinder Fill the cylinder to 10 ml with clean solvent...

Страница 11: ...m Pour 5 ml of the 50 ppm standard back into the 10 ml graduated cylinder Fill the cylinder to 10 ml with clean solvent This will cut the 50 ppm standard in half to be 25 ppm Pour this standard into a...

Страница 12: ...Calibration to Prepared Standards Make sure InfraCal has been on for 1 hour Make sure cuvette is clean Make sure cuvette is clean Zero InfraCal...

Страница 13: ...Zeroing Fill the cuvette with clean solvent and place in the sample stage with the frosted side facing front...

Страница 14: ...Zeroing Press and hold the ZERO button until the display reads bAL Recall ZERO CAL RUN bAL...

Страница 15: ...Zeroing A multiplier value to 3 decimal places will be displayed when zero is established The actual value is only of interest when reporting problems to the factory Recall ZERO CAL RUN 1 267...

Страница 16: ...Zeroing Press RUN The display should read 00 02 If not repeat the zero process Recall ZERO CAL RUN 00...

Страница 17: ...red Standards Run each standard at least 3 times on the InfraCal Average results for each standard Enter averaged results under the A in the manual on page 13 or 14 and the corresponding calibration v...

Страница 18: ...Example Calibration Table A Absorbance C Calibration Standard AO1 42 CO1 25 AO2 81 CO2 50 AO3 149 CO3 100...

Страница 19: ...ntering Values into EDIT program Entering Values into EDIT program Make sure Cal Lock switch on the back of the InfraCal is in the I position Made In U S A EN 61010 1 EN 55022 Class B EN 50082 1 Seria...

Страница 20: ...Press the CAL button for two seconds until CAL appears on the display CAL Recall ZERO CAL RUN...

Страница 21: ...Press RECALL until Edit is displayed Edit Recall ZERO CAL RUN...

Страница 22: ...3 Press CAL and the display will read n for a short time n 3 Recall ZERO CAL RUN n...

Страница 23: ...3 Set n to the number standards using the up and down arrows n 3 Recall ZERO CAL RUN n...

Страница 24: ...3 Set n to the number standards using the up and down arrows for our example it will be 3 3 3 Recall ZERO CAL RUN 3...

Страница 25: ...Press CAL and the display will read A01 for a short time A01 Recall ZERO CAL RUN A01...

Страница 26: ...Absorbance Calibration Standard AO1 42 CO1 25 Using the up and down arrows scroll to the first absorbance value for our example it will be 42 AO1 42 CO1 25 AO2 81 CO2 50 AO3 149 CO3 100...

Страница 27: ...42 Using the up and down arrows scroll to the first absorbance value for our example it will be 42 Recall ZERO CAL RUN 42...

Страница 28: ...Press CAL again and the display will read C01 C01 Recall ZERO CAL RUN C01...

Страница 29: ...Absorbance Calibration Standard AO1 42 CO1 25 Using the up and down arrows scroll to the first calibration value for our example it will be 25 CO1 25 AO2 81 CO2 50 AO3 149 CO3 100...

Страница 30: ...25 Using the up and down arrows scroll to the first calibration value for our example it will be 25 Recall ZERO CAL RUN 25...

Страница 31: ...play will read A02 for a short time Using the up and down arrows scroll to the first calibration value for our example it will be 81 Absorbance Calibration Standard AO1 42 CO1 25 AO1 42 CO1 25 AO2 81...

Страница 32: ...Press CAL again and the display will read C02 Scroll this to the second calibration value Absorbance Calibration Standard AO1 42 CO1 25 AO2 81 CO2 50 AO3 149 CO3 100...

Страница 33: ...Do the same for AO3 and CO3 Absorbance Calibration Standard AO1 42 CO1 25 AO2 81 CO2 50 AO3 149 CO3 100...

Страница 34: ...Press CAL again and the display will read IDLE IDLE Recall ZERO CAL RUN The calibration is now stored and the calibration is now active in the InfraCal...

Страница 35: ...InfraCal TOG TPH Analyzer Model CVH Calibration For questions please contact us at tech WilksIR com or 203 855 9136...

Отзывы: