Page 13 of 15 

 

Test points 

7.0      Reference test points

 

 
Test points located in the battery compartment allow access to the control voltage used as part of 
the motor feedback loop. Using an accurate AC DMM and/or an oscilloscope can help make a 
determination as to the condition of the unit. 
 
To access the reference test points, remove the battery cover (see section 4) and find the two turret 
style connectors and connect the desired test leads.  

The “+” terminal is the AC output voltage. The  

“-“ terminal is system ground. 

 

CAUTION

:  When accessing the test points, proper precautions should be taken against ESD 

(electrostatic discharge).  Checking the test point voltage should be done only in an ESD-
safe environment.  

 

 
 

 
 
 
 
 
 
 
 
 

Figure 5:  Test points  

 

7.1 

 Background information

 

 

The REF2500 contains an internal sine wave generator with a reference accelerometer in the 
control loop to maintain a continuous output level of 1g in either peak or RMS.  The reference 
accelerometer’s signal is fed through a charge amplifier to the control electronics.

 

 
Proper operation of the feedback control loop and accuracy of the internal voltage reference can be 
verified by monitoring the voltage at the test point output.  Since we are observing the accelerometer 
signal, the test points will also provide information as to any mechanical or significant electrical 
anomalies. 
  
In the event the REF2500 is dropped or otherwise abused, measuring the output from the test points 
will confirm if the unit is still operational. 

 
 
 
 
 
 
 

Содержание REF2500

Страница 1: ...rive Frederick MD 21701 USA Amphenol Maryland Inc d b a Wilcoxon Sensing Technologies 97029 Rev C 1 09 18 Tel 1 301 330 8811 Tel 1 800 WILCOXON Fax 1 301 330 8873 www wilcoxon com ReferenceMate portab...

Страница 2: ...uld be consulted and read carefully CAUTION This symbol is used when caution is needed to prevent damage to equipment It is used where careful attention to certain procedures described in the instruct...

Страница 3: ...ower ON OFF 9 5 2 3 Frequency selection 10 5 2 4 RMS PEAK 10 5 2 5 BAT low power LED 11 5 2 6 OOR LED 11 5 2 7Orientation during operation 11 6 0 Testing triaxial transducers 12 6 1 Universal AC adapt...

Страница 4: ...z for imperial measurements where 1 g is equal to 1 inch per second ips ideal for use with velocity sensors 100 0 Hz to confirm accelerometer performance at the same reference frequency used by most m...

Страница 5: ...Remove the battery holder and insert four AA batteries into the holder being careful to observe correct polarity see figure 1B After the batteries are installed insert the holder into the battery cup...

Страница 6: ...and any mounting accessory is mounted to the ReferenceMate head connect all cables from the UUT to the readout equipment You are now ready to test your device 3 Press the ON OFF button to initiate the...

Страница 7: ...n the ReferenceMate off before the end of its autoshutdown cycle Press the ON OFF button Figure 2 control panel of the ReferenceMate 5 2 Operational details Peak RMS mode indicators Indicate peak or R...

Страница 8: ...n RMS The equivalent peak to peak displacement values in mils are shown in Table 1a for reference For those using the metric system Table 1b shows the equivalent values All the values below are repres...

Страница 9: ...712 sensors 24 20 2 7 2 3 13249 02 1 4 28 to 10 32 adaptor plate 24 18 2 7 2 0 13267 01 90 adaptor plate 993 series sensors 24 24 2 7 2 7 SF6M 1 1 4 28 to M6 adaptor stud 24 30 2 7 3 4 SF6M 1 4 28 to...

Страница 10: ...l units US One g at 61 4 Hz is equal to 1 in sec ips velocity 100 Hz is the preferred frequency for checking accelerometer performance against calibration data 100 Hz is the most common reference freq...

Страница 11: ...ed OOR LED will be illuminated during the control time cycle Once the LED goes out the unit is ready for operation This is normally less than 3 seconds 5 2 7 Orientation during operation The Reference...

Страница 12: ...eration of the ReferenceMate2500 is required part number REF003 is available This AC adaptor comes with several universal plug adaptors and will operate from 50 60 Hz 110 220 VAC To power the shaker w...

Страница 13: ...the test point voltage should be done only in an ESD safe environment Figure 5 Test points 7 1 Background information The REF2500 contains an internal sine wave generator with a reference acceleromet...

Страница 14: ...f high levels of distortion an incorrect output level or out of tolerance frequency is observed the REF2500 should be returned to the factory for service There are no user serviceable parts in the uni...

Страница 15: ...ack cover must be securely installed on the unit Do not expose the unit to dust or moisture without both of these protective mechanisms in place 8 0 Storage If the unit is going to be out of service f...

Отзывы: