EN
11258421.13 10/2017 EN/DE
WIKA operating instructions model T32.xS
20
6. Commissioning, operation
WIKA T32 configuration software
We recommend using our WIKA T32 configuration software. This software is regularly
updated and adapted to the firmware extensions of the T32, so that you always have full
access to all functionalities and parameters of the transmitter (see chapter 8 “WIKA T32
configuration software”).
Further configuration software
With the following software tools it is also possible to carry out configurations at the T32 e.g.:
■
AMS and SIMATIC PDM (T32_EDD)
■
FieldMate, PACTware, SmartVision and Fieldcare (DTM_T32)
■
DTM in FDT 1.2 frame application
With any other HART
®
configuration tool the generic mode functionalities can be operated
(e.g. measuring range or Tag No.).
Further information on the configuration of the T32 with the software tools
mentioned above is available on request.
6.3.2 DD version
The model T32.xS temperature transmitter can be used with the following DTM and DD
versions.
T32 HART
®
instrument
revision
Corresponding DD
(Device Description)
T32 HART
®
DTM
0
Dev v0, DD v2
DTM 1.0.2
1
Dev v1, DD v1
DTM 1.0.2
2
Dev v2, DD v1
DTM 1.0.2
3
Dev v3, DD v1
DTM 2.0.0.175, DTM 2.1.0
Optional: HART
®
7 version
4
Dev v3, DD v1
DTM 2.1.0
6.3.3 HART
®
communicator (FC375, FC475, MFC4150, MFC5150)
With the HART
®
communicator the instrument functions are selected via various menu
levels and with the help of a special HART
®
function matrix (see chapter 6.5 “HART
®
configuration tree”).
6.4 Connection of FSK modem, HART
®
communicator
WARNING!
■
The measuring circuit has to have a load of at least 250 Ω.
■
For all transmitters with hazardous area protection observe chapter 10
“Notes for mounting and operating in hazardous areas”.
This resistor is already integrated in most power supplies available in the market and is
therefore not required separately. Frequently a special connection for the FSK modem is
already available.
Содержание T32.1S
Страница 38: ...EN 11258421 13 10 2017 EN DE WIKA operating instructions model T32 xS 38 ...
Страница 74: ...DE WIKA Betriebsanleitung Typ T32 xS 74 11258421 13 10 2017 EN DE Anlage 1 Installation drawing CSA FM ...
Страница 75: ...DE WIKA Betriebsanleitung Typ T32 xS 75 11258421 13 10 2017 EN DE Anlage 1 Installation drawing CSA FM ...
Страница 76: ...DE WIKA Betriebsanleitung Typ T32 xS 76 11258421 13 10 2017 EN DE Anlage 1 Installation drawing CSA FM ...
Страница 77: ...DE WIKA Betriebsanleitung Typ T32 xS 77 11258421 13 10 2017 EN DE Anlage 1 Installation drawing CSA FM ...
Страница 78: ...DE WIKA Betriebsanleitung Typ T32 xS 78 11258421 13 10 2017 EN DE Anlage 2 EU Konformitätserklärung ...
Страница 79: ...11258421 13 10 2017 EN DE WIKA operating instructions model T32 xS 79 ...