WIKA operating instructions flow switch, model FSD-3
EN
14153887.01 09/2016 EN/DE/FR
Specifications
Voltage supply
Power supply
DC 15 ... 35 V
Current consumption
■
Switching output with analogue signal: 175 mA
■
Switching output without analogue signal: 150 mA
Total current consumption
max. 650 mA including switching current
Accuracy specifications
Non-repeatability
Flow (5 ... 100 cm/s): ±2 cm/s
Temperature: 0.5 K
Accuracy at reference conditions
Flow (5 ... 100 cm/s): ≤ ±5 % of end value of measuring range
Flow (>100 ... 175 cm/s): ≤ ±10 % of end value of measuring range
Temperature: ≤ ±1,5 K
Including non-linearity, hysteresis, zero offset and end value deviation (corresponds to
measured error per IEC 61298-2).
The accuracy of the flow switch is dependent on the thermal conductivity and the contamination
of the probe.
Temperature error at -20 ... +85 °C
(-4 ... +185 °F)
≤ ±0.4 cm/s per K
Reference conditions
Ambient temperature
15 ... 25 °C (59 ... 77 °F)
Atmospheric pressure
860 ... 1,060 mbar (12.47 ... 15.38 psi)
Humidity
45 ... 75 % r. h.
Medium
Water
Power supply
DC 24 V
Mounting position
Process connection M18 x 1.5 downwards
Inner diameter of pipe 26 mm
Upstream/Downstream pipe: 1 m / 0.5 m
Marking towards the upstream side twist of ±5°
Load
100 Ω
9. Specifications
Содержание FSD-3
Страница 38: ...38 WIKA operating instructions flow switch model FSD 3 EN 14153887 01 09 2016 EN DE FR...
Страница 74: ...74 WIKA Betriebsanleitung Str mungsschalter Typ FSD 3 DE 14153887 01 09 2016 EN DE FR...
Страница 110: ...110 WIKA mode d emploi pour le capteur de d bit type FSD 3 FR 14153887 01 09 2016 EN DE FR...
Страница 111: ...111 WIKA mode d emploi pour le capteur de d bit type FSD 3 FR 14153887 01 09 2016 EN DE FR...