5.4 Electrical connection
To prevent interferences from coupling into the system, observe the following
information:
■
A shielded, low-capacitance measuring cable is attached to the single point load cell.
■
Ground the shield of the measuring cable.
■
Avoid stray fields of transformers, motors and contactors.
■
When using extensions, only shielded and low-capacitance cables should be used.
The permitted maximum and minimum lengths of cable are defined in ISO 11898-2.
Care should be taken also to ensure a high-quality connection of the shielding.
■
Do not install measuring cables in parallel to 3-phase-current cables and control
cables.
■
Single point load cell, amplifiers and processing or display units must not be ground-
ed several times.
The pin assignment of the cable can be found on the product label.
Exc+
Sig+
Exc-
Sig-
UB+ (red)
S+ (green)
UB- (black)
S- (white)
(Shield)
Electrical connection
Supply voltage (+)
UB+ or E+
Red
Supply voltage (-)
UB- or E-
Black
Signal (+)
S+
Green
Signal (-)
S-
White
Shield
Shield
Shield
5.4.1 Cable pin assignment
5. Commissioning, operation