2. Safety
WIKA Operating Instruction, CPP4000-X
9
EN
14
15
54
60
.0
08
/20
16
GB
2.6.1
Mineral oils health and safety information
DH-Budenberg/WIKA provide hydraulic mineral oil in containers up to 500 ml, labelled “ISO VG 22”,
for use up to 4,000 bar in test pumps. It is no more hazardous than other common lubricating oils.
Information
The term “reference measuring instrument” in these operating instructions refers to
any pressure measuring instrument, such as: pressure gauge, electrical pressure
measuring instrument and pressure transmitter with electrical output. The hydraulic
comparison test pump system is only as accurate as the reference measuring
instrument used. The reference measuring instrument should be regularly calibrated
in order to ensure that its accuracy is maintained.
Description
Closed flash point
greater than 120 °C
Storage
not above 30 °C
Oral LD 50
15 g per kg body weight
Threshold limit value
5 mg/m³
Fire extinguishing media
CO2/dry chemical foam or water fog
Spillage
Soak with absorbent clay or proprietary absorbent
Waste disposal
Burn or dump in approved area
Emergency treatment of acute effects
Ingestion
Do not induce vomiting.
Administer 250 ml milk or olive oil.
The main hazard following accidental ingestion is aspiration of liquid
into lungs.
Aspiration
Send to hospital immediately
Inhalation
Remove to fresh air, if nausea persists seek medical attention.
Eye contact
Wash with copious amounts of water for at least 10 minutes.
If irritation results or persists, obtain medical advice.
Skin contact
Where skin rashes or other abnormalities occur as a result of
prolonged or repeated contact, medical advice should be obtained as
soon as possible.
2.6.2
Other liquids
For some very particular applications we supply specially constructed liquids. Copies of
manufacturer’s data can be sent on request.
Information
If the hydraulic comparison test pump is used with water as the pressure medium, for
pressures greater than 700 bar, it is recommended that a mixture of 90 % water and
10 % glycerine (by volume) is used.