27
WIKA operating instructions, model CPH6400
11313439.04 12/2012 GB/D
GB
6. Commissioning, operation
6.4.4 Info screen
Via the
MODE
key different functions can be superimposed, e.g. the function: Info (
INFO
).
Activation of the Info screen
Pressing the MODE key four times* (directly within the standard measuring mode)
* If you are in a different measuring mode than the standard measuring mode, press the
MODE
key until the
measuring mode with the
Info Screen
function appears on the display).
1) Serial number of the CPH6400
2) Current Firmware version in the instrument
3) Internal sensor number of the currently-connected sensor
4) Pressure type for the currently-connected sensor
5) Accuracy in % FS of the currently-connected sensor
6) Lower limit of the measuring range for the currently-connected sensor
7) Upper limit of the measuring range for the currently-connected sensor
8) Calibration date of the currently-connected sensor
MODE
Return to the standard measuring mode by pressing the MODE key several
times.
6.5 Pressure units
Via the
UNIT
key, depending on the measuring range of the current pressure sensor, the
units bar, mbar, kPa, psi, mmHg, inHg or kg/cm² can be selected.
Press the
UNIT
key until the required unit is displayed (ring-shift).
6.6 Measurement signal processing
The standard value for the filter for the measuring signal processing/smoothing is a factor
of "3" and can only be altered, to a value between 1 and 5, using the WIKA "PrintCal" PC
software.
1 = no filter, ... , 5 = high filter
Info screen
1
3
5
7
2
4
6
8
Содержание CPH6400
Страница 40: ...40 WIKA operating instructions model CPH6400 11313439 04 12 2012 GB D GB ...
Страница 76: ...11313439 04 12 2012 GB D 76 WIKA Betriebsanleitung Typ CPH6400 D Anlage 1 Konformitätserklärung Typ CPH6400 ...
Страница 77: ...11313439 04 12 2012 GB D 77 WIKA Betriebsanleitung Typ CPH6400 D Anlage 2 Konformitätserklärung Typ CPT6400 ...