3. Specifications
3. Specifications
Specifications
Precision hand-held pressure indicator
model CPH6400 (complete measuring chain)
Sensor technology
1 reference pressure sensor (interchangeable without tools)
1)
Measuring range
bar
0.4
1.6
6
16
40
100
250
600
1000
Overpressure safety
bar
2
10
35
80
80
200
500
1200 1500
Burst pressure
bar
2.4
12
42
96
400
800
1200 2400 3000
Accuracy of the measuring chain
0.025 % FS
2)
Measuring range
bar
1600 2500 4000 5000 6000
Overpressure safety
bar
2300 3500 5000 6000 7000
Burst pressure
bar
4000 6000 8000 10000 11000
Accuracy of the measuring chain
0.1 % FS
2)
Type of pressure
{In addition to the above-specified pressures; vacuum, bi-direction-
al ranges and absolute pressures are available}
Pressure units
bar, mbar, kPa, psi, mmHg, inHg and kg/cm² (depending on the
measuring range, selectable)
Active temperature compensation
°C
10 ... 40
Temperature measurement
only for 2-channel version
■
Sensor type
Pt100, 4-wire
■
Measuring range
°C
-10 ... +50
■
Resolution
°C
0.01
■
Accuracy
°C
0.05
Permissible ambient temperature
°C
0 ... 50
Calibration
2)
Factory calibration certificate (optional: DKD/DAkkS calibration
certificate)
Specifications
Reference pressure sensor model CPT6400
Pressure connection
≤ 1000 bar: G ½ B; {various connection adapters on request}
> 1000 bar: M16 x 1.5 female, with sealing cone
Material
Stainless steel for wetted parts
(for measuring ranges > 25 bar ... ≤ 1000 bar additional Elgiloy
®
)
Internal transmission fluid
Synthetic oil (only for measuring ranges up to 25 bar)
Permissible temperature ranges
■
Medium
°C
-20 ... +80
■
Storage
°C
-40 ... +85
Case
Stainless steel
Electrical connection
Circular connector, 8-pin
Ingress protection
IP 65 (with cable connected)
Weight
g
approx. 220
CE conformity
■
Pressure equipment directive
97/23/EC, PS > 200 bar; module A, pressure accessory
■
EMC directive
2004/108/EC, EN 61326 Emission (Group 1, Class B) and Immuni-
ty (industrial locations)
1) Calibration data sets for up to 10 transmitters can be stored in each instrument (up to 10 calibration data sets)
2) Calibrated at 23 °C and in vertical mounting position with pressure connection facing downwards.
{} Items in curved brackets are optional extras for an additional price.
Содержание CPH6400
Страница 40: ...40 WIKA operating instructions model CPH6400 11313439 04 12 2012 GB D GB ...
Страница 76: ...11313439 04 12 2012 GB D 76 WIKA Betriebsanleitung Typ CPH6400 D Anlage 1 Konformitätserklärung Typ CPH6400 ...
Страница 77: ...11313439 04 12 2012 GB D 77 WIKA Betriebsanleitung Typ CPH6400 D Anlage 2 Konformitätserklärung Typ CPT6400 ...