56
WIKA operating instructions ProcessCalibrator, model CPH6000
11069023.04 11/2012 GB/D
GB
..Ref. Sensor-list
362A
R.-Start:
Range-End:
Unit:
Cal-Dat.:
Class:
0.000
10.000
bar
2011/05/10
0.025
Rp56
362A
Kd35
----
----
----
----
----
01
02
03
04
05
06
07
08
09
10
Rp52
----
Current
a
b
c
g) Ref-Value:
Current measured value of the connected CPT6000 reference pressure sensor
h) Offset:
This menu option only appears if the CPH6000’s reference pressure sensor is an
absolute pressure sensor.
Through this menu option the measured value of the reference pressure sensor
can be adjusted. This should only be used, however, as close as possible to the
absolute zero, and using a reference that is at least 4 times more accurate.
i) Cal-Dat.:
Calibration date of the CPT6000 reference pressure sensor (Year/Month/Day)
7.4.4 SETUP additional menu items: Ref. Sensor-list
7. Commissioning, operation
The ProcessCalibrator supports up to 10 CPT6000 reference pressure sensors. These
are listed in this menu.
a) Currently connected CPT6000 reference pressure sensor
b) Sensor list of the supported CPT6000 reference pressure sensors (calibrated with
the instrument)
c) Date of the sensor which has been selected using the cursor (Calibration date:
Year/Month/Day)
Selection
of a listed sensor