WIKA operating instructions ProcessCalibrator, model CPH6000
11069023.04 11/2012 GB/D
GB
a
b
c
d
e
f
7. Commissioning, operation
a) Measuring range of the CPT6000 reference pressure sensor (which is currently
connected)
b) Current measured value of the reference pressure sensor
c) Pressure unit (settable via menu)
In
MEASURING
mode, at the same time as the reference-pressure value, a test
item can also be displayed on the screen (see following picture). For configuration
instructions, please see chapters 7.3.1 "MEASURING mode" and 7.3.2 "MEASURING
mode (with test item)".
Display:
MEASURING
with test item
a) Measuring range of the test item
b) Current measured value of the test item
c) Deviation/difference between reference and test item in the current pressure units
and in % of the measuring span (% FS) or % of measured value (% rd)
d) Pressure unit (of the test item)
e) Original output signal of the test item
f) Current value of the output signal of the test item
Pressure signal
(test item)
Electrical signal
(test item)
P
I/U
switchable