Current cursor position; alter via
◁▷ ◀▶
Parameter selection from list or menu via
0.00
Parameter input via numeric keypad
ENTER
Input confirmation
CLEAR
Clear entry
Customer-specific unit; with respect to bar
(input via numeric keypad)
Display resolution in operating mode
(back with
)
Select and confirm (standard units) via
If the instrument is switched into
"MEASURING"
mode (with test item = display the
test signal as an electrical signal or as a pressure), in order to carry out a comparative
measurement or calibration without measured-value recording, then follow the
instructions on the previous page.
The following is a more detailed explanation of point "2. Preparation for
MEASURING"
a) Test item type and respective test item measuring signal
[0 ... 20 mA / 4 ... 20 mA / 0 ...
1 V / 0 ... 2 V / 0 ... 5 V / 0 ... 10 V / or mechanical for dial pressure gauges]
If a comparative measurement with a mechanical gauge (test item) is being made, then
the gauge’s measured value should be entered via the numeric keypad and confirmed
with the
ENTER
key.
b) R.-Start/Range-End:
Start of measuring range and end of measuring range of the currently to calibrate test
item
c) Unit and resolution (submenu)
7. Commissioning, operation
d) Measurement uncertainty of the test item in % FS (i.e. of the span) or % rd (i.e. of the
measured value)
e) Measurement type for the test item (relative or absolute)
f) Test medium (pneumatic -> gas or hydraulic -> oil)
g) Voltage supply for test item (on/off)
[If no external supply is required for the test
item, "OFF" should be selected to conserve energy]