background image

INSTRUCTION MANUAL 

NI-205WE 

Rev. 3 12/16

 

 

4 of 8 

Use  a  wide  bladed  screwdriver,  as  indicated  on  the 
adhesive plate, until the relative lamp turns on (or turns 
off). 
- If the instrument is equipped with only one contact the 
calibration is complete. 
-  If  it  is  equipped  with  two  contacts  continue  in  the 
following manner. 
Vary the pressure until the desired set point value for the 
second microswitch is reached (fig. 3). 
Act on the adjustment screw of the second contact. 
Repeat  former  operations  on  the  first  contact,  then 
operations  on the second contact, until the  required set 
point precision is obtained. This is necessary due to the 
reciprocal influence which the microswitches have on the 
sensitive element of the instrument. 

CAUTION: 

if the two set point are different they must be 

different for much of 5% of the adjustable span. 

6.3  CHECK OF SET POINT 

Generate  the  normal  working  pressure  and  wait  the 
pressure  stabilisation.  Vary  the  pressure  into  the  circuit 
and record the set point value. Write the set point values 
on the adhesive label. 

Note

:  the  repeatability  should  be  checked  verifying  for 

three  times  the  set  point  (Pi)  starting  always  from  the 
same pressure value (Pw). The pressure cycle should be 
slowly to give the possibility to record the set point with 
accuracy.. 

6.4  CALIBRATION  CIRCUIT  FOR  INSTRUMENT 

WITH ADJUSTABLE RANGE LESS TO 60 mbar 

The  calibration  circuit  used  for  the  calibration  of  these 
instruments must be: 
-  of  big  internal  volume  (5  liters  or  bigger)  in  order  to 
reduce the effect of volumetric variation (and therefore of 
pressure) caused by the sensing element of the pressure 
switch during the snap action. 
- in a thermally stable place in order to guarantee stability 
to the pressure inside the circuit used for the calibration. 
Have  to  be  considered  that  in  a  closed  circuit  with  the 
internal pressure equal to the atmospheric pressure, the 
variation  of  temperature  of  1°C  causes  in  the  circuit  a 
pressure variation of 3,4 mbar. 
The maximum inlet pressure must not exceed the lesser 
of overpressure allowed by the pressure switch and the 
test gauge. 
The  test  instrument  should  have  a  measurement  range 
approximately equal to or slightly wider than the pressure 
switch range and should have an accuracy consistent with 
the accuracy required to calibrate the set point. 
For example APW10 range 0..16mbar the accuracy of the 
test gauge must be ± 0,04 mbar to calibrate the set point 
with  an  accuracy  of  ±  0,16  mbar  (1%  of  the  adjustable 
span).

 

Fig.  6  –  Calibration  circuit  for  instrument  with  adjustable 
range less than 60 mbar 

 

1-  Pressure switch 
2-  Test gauge 
3-  Inlet valve 

4-  Volumetric pump  
5-  Capacity 
6-  Outlet valve 

6.5  ADJUSTMENT  OF  DEAD  BAND  (LETTER  R  ON 

THE MODEL CODE) 

The dead band can be adjusted only on the instrument is 
equipped  with  a  microswitch,  which  allows  adjusting 
(Letter  R  on  the  model  code).  Adjustment  may  be 
obtained by rotating the wheel placed on the microswitch 
(Fig. 7). 
In order to carry out this operation it is advisable to use a 
blade screwdriver. 

Fig. 7 – Adjustment of dead band 

 

  Set point calibration screw 

  Dead band adjustment wheel 

WARNING:

  rotate  the  wheel  without  pushing  too  much 

the blade against it. 
The  instrument  is  normally  delivered  adjusted  on  the 

minmum value

 of its range (factory calibration). 

Calibration of dead band 

The  calibration  of  the  dead  band  is  obtained  using  the 
following procedure: 
1 -  Raise  pressure  in  the  circuit  until  reaching  the  set 

point and record its value (Pi). 

2 -  Reduce pressure in the circuit until reaching the reset 

point and record its value (Pr). 

3 -  The difference Pi - Pr = Va represents the dead band 

factory adjusted value. 

4 -  Rotate the adjustment wheel in the sense shown in 

Fig. 6 placing the red notch in horizontal position 

5 -  Repeat operations 1 and 2 and measure the new dead 

band Vb. 

6 -  By  comparing  the  values  Va  and  Vb  approximately 

determine the color of the wheel notch to be placed on 
the mark. 

7 -  Place the notch and measure the obtained dead band. 
8 -  Proceed by successive approximations until reaching 

the desired dead band value with enough accuracy. 

9 -  Then proceed with the set point calibration 

Example:

 The dead band increase corresponding to the 

rotation from A to B is given by: Vb - Va =I

 

The  desired  dead  band  V  will  be  approximately  in  the 
position  indicated  by  the  value 

I

V

K

    which 

expresses: 

By  units,  the  wheel  notches  (1=one  black  notch, 
2=one  red  notch,  3=one  yellow  notch,  4=one  blue 
notch, 5=one green notch, 6=two black notches). 

By  decimal  digits,  the  percentage  middle  position 
between the located notch (of units) and the following 
one. 

6.6  FINAL OPERATIONS 

Disconnect the instrument from the calibration circuit. 

6.6.1 Weatherproof pressure switches (Models APW) 

Take the cover, ensure that the sealing gasket is correctly 
fitted into its seat, and insert the cover onto the case, with 
the  blocking  gap  positioned  in  correspondence  to  the 
blocking  bracket.  Turn  the  cover  clockwise  closing  it 
tightly. Mount the adjustment screw access plate, then the 
blocking device as in Fig.4. 

Содержание APW SERIES

Страница 1: ...ions in mm Versions APW10 Versions APA10 A pressure attachment B cable entry WEIGHT 8 2 kg dimensions in mm A vacuum attachment B cable entry WEIGHT 10 kg dimensions in mm For surface mounting use fou...

Страница 2: ...otection against heat radiation fluid separators cooling coils heated lockers must be taken Process fluid or its solid particles must not however go inside instrument chambers 2 OPERATING PRINCIPLE Th...

Страница 3: ...1 Weatherproof pressure switches Models APW Remove the blocking device fixed to the side of the instrument case and the adjustment screw access plate fig 59 Remove the cover by rotating it in an anti...

Страница 4: ...brate the set point with an accuracy of 0 16 mbar 1 of the adjustable span Fig 6 Calibration circuit for instrument with adjustable range less than 60 mbar 1 Pressure switch 2 Test gauge 3 Inlet valve...

Страница 5: ...to the applicable standards In case of flameproof instrument see also the Standard IEC 60079 14 If the electrical connection is carried out in a protected tube it shall be made so that condensate is...

Страница 6: ...ere should be no trace of leakage of process fluid outside the instrument In case of flame proof or intrinsic safety instruments inspections of the electrical installation are to be carried out also a...

Страница 7: ...TION PROBABLE CAUSE REMEDY Set point shift Permanent deformation of the sensitive element due to fatigue or non tolerated overranges Variation of the elastic features of the sensitive element due to i...

Страница 8: ...fitting 10 Three piece fitting 11 Curve 12 Blocking joint 13 M6 screws No 4 14 Bracket for 2 pipe 15 Vertical pipe 16 Horizontal pipe Annex 1 Model code 1 MODEL CODE AP For further information see dat...

Отзывы: